STIFF 25G TANO DIAMOND DUSTED MEMBRANE SCRAPER
Report
- Report Number
- 0001932402-2023-70003
- Event Type
- Injury
- Date Received
- July 26, 2023
- Date of Event
- June 15, 2023
- Report Date
- June 19, 2023
- Manufacturer
- SYNERGETICS
- Product Code
- GZX
- UDI-DI
- 20841305108848
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT WAS RECEIVED FOR EVALUATION OPEN IN ITS ORIGINAL POUCH. THE TIP PROTECTOR WAS ON THE END OF THE HANDLE. THE TUBE AND INTERNAL CARBIDE ROD THAT MAKE UP THE SHAFT HAD BROKEN OFF APPROXIMATELY 2MM FROM THE HANDLE AND WAS TAPED TO A YELLOW PIECE OF PAPER. THE TUBE AT THE POINT OF THE BREAK APPEARED STRETCHED AND TORN. THE CARBIDE IS CLEANLY BROKEN. THE PROBE APPEARS TO HAVE BEEN EXPOSED TO RADIAL STRESSES THAT EXCEEDED THE STRENGTH OF THE SHAFT. INFORMATION SUPPLIED BY USER CONFIRMED, DEVICE WAS USED WITH A VALVED CANNULA. INSTRUCTIONS FOR USE WARNINGS STATE, ONLY RETRACTABLE DDMS PRODUCTS ARE COMPATIBLE WITH VALVED CANNULAS. MANUFACTURING AND STERILIZATION RECORDS WERE REVIEWED AND FOUND TO BE ACCEPTABLE. THE LOT HISTORY, TREND ANALYSIS, RISK ANALYSIS AND/OR DIRECTIONS FOR USE REVIEW WERE CONSIDERED ACCEPTABLE, WITH THE PRODUCT PERFORMING WITHIN ANTICIPATED RATES. NO CORRECTIVE ACTION REQUIRED.
IT WAS REPORTED THAT THE TIP OF THE DIAMOND DUSTED MEMBRANE SCRAPER BROKE OFF AT THE HUB INTO THE BACK OF THE PATIENT''S EYE AND INJURED THE RETINA (AFTER IT BROKE OFF OF HANDLE) CAUSING MICROSCOPIC TEARING. THE SURGEON REMOVED THE TIP FROM THE EYE, AND LASERED THE BACK OF THE EYE WHERE THE DIAMOND DUSTED TIP HIT. THE SURGERY WAS PROLONGED AN ADDITIONAL THREE MINUTES. THE PATIENT IS DOING WELL AFTER SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1840715 | STIFF 25G TANO DIAMOND DUSTED MEMBRANE SCRAPER | INSTRUMENT, MICROSURGICAL | GZX | SYNERGETICS | 20.04.25S | M0043153 | 20841305108848 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | STELLARIS UNIT AND ACCESSORIES |