FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 17396001 · Received July 26, 2023

Report

Report Number
1000113657-2023-00391
Event Type
Malfunction
Date Received
July 26, 2023
Date of Event
June 29, 2023
Report Date
August 9, 2023
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 09-AUG-2023: H3: DEVICE EVALUATED BY MANUFACTURER. H6: UPDATED FDA'S TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS. H10: METER AND TEST STRIPS WERE RETURNED- DEFECT FOUND ON RETURNED STRIPS: HIGH RESULTS. REPORTED DEFECT NOT REPRODUCED ON RETURNED METER. PRODUCT EVALUATION HAS BEEN COMPLETED. R&D INVESTIGATION HAS BEEN COMPLETED AND ROOT CAUSE SELECTED. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIPS TESTED WITHIN SPECIFICATIONS. ROOT CAUSE: RC-072: VIAL LEFT OPEN FOR AN EXTENDED PERIOD OF TIME.

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE RETURNED - PRODUCT EVALUATION IN-PROCESS. NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 198, 268, 212, 224 AND 222 MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100-160 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE OFFICE. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 10/31/2024 AND OPEN VIAL DATE IS 06/2023. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 198 MG/DL DATE: 6/29/23 TIME: 9:46 AM FASTING RESULT 2: 268 MG/DL DATE: 6/28/23 TIME: 6:49 AM FASTING RESULT 3: 212 MG/DL DATE: 6/27/23 TIME: 6:18 AM FASTING RESULT 4: 224 MG/DL DATE: 6/27/23 TIME: 6:18 AM FASTING RESULT 5: 222 MG/DL DATE: 6/26/23 TIME: 8:29 AM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2184803 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TMX WGN 100CT12/CS MG/DL #383569 ZB5228S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown