TRUE METRIX
Report
- Report Number
- 1000113657-2023-00391
- Event Type
- Malfunction
- Date Received
- July 26, 2023
- Date of Event
- June 29, 2023
- Report Date
- August 9, 2023
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
SECTIONS WITH ADDITIONAL INFORMATION AS OF 09-AUG-2023: H3: DEVICE EVALUATED BY MANUFACTURER. H6: UPDATED FDA'S TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS. H10: METER AND TEST STRIPS WERE RETURNED- DEFECT FOUND ON RETURNED STRIPS: HIGH RESULTS. REPORTED DEFECT NOT REPRODUCED ON RETURNED METER. PRODUCT EVALUATION HAS BEEN COMPLETED. R&D INVESTIGATION HAS BEEN COMPLETED AND ROOT CAUSE SELECTED. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIPS TESTED WITHIN SPECIFICATIONS. ROOT CAUSE: RC-072: VIAL LEFT OPEN FOR AN EXTENDED PERIOD OF TIME.
INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE RETURNED - PRODUCT EVALUATION IN-PROCESS. NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 198, 268, 212, 224 AND 222 MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100-160 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE OFFICE. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 10/31/2024 AND OPEN VIAL DATE IS 06/2023. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 198 MG/DL DATE: 6/29/23 TIME: 9:46 AM FASTING RESULT 2: 268 MG/DL DATE: 6/28/23 TIME: 6:49 AM FASTING RESULT 3: 212 MG/DL DATE: 6/27/23 TIME: 6:18 AM FASTING RESULT 4: 224 MG/DL DATE: 6/27/23 TIME: 6:18 AM FASTING RESULT 5: 222 MG/DL DATE: 6/26/23 TIME: 8:29 AM FASTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2184803 | TRUE METRIX | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TMX WGN 100CT12/CS MG/DL #383569 | ZB5228S | 00021292007836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |