FDA Adverse Event
Injury
Summary report: N
RAD CAP RHEAD RADIAL HEAD AND STEM
MDR report key: 1739577
·
Received June 25, 2010
Report
- Report Number
- 3003640913-2010-00021
- Event Type
- Injury
- Date Received
- June 25, 2010
- Date of Event
- May 1, 2010
- Report Date
- June 25, 2010
- Manufacturer
- SMALL BONE INNOVATIONS, INC.
- Product Code
- JDB
- PMA / PMN Number
- K060038
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DIMENSIONAL EVALUATION SHOWED DEVICES WITHIN SPECIFICATION. VISUAL EXAMINATION SHOWED DAMAGE TO STEM COMPONENT THAT WOULD PREVENT HEAD FROM FULLY ENGAGING. THIS DAMAGE IS CONSISTENT WITH HEAVY IMPACT ON STEM AS WITH METAL HAMMER.
Description of Event or Problem · 1
PATIENT HAD A RADIAL HEAD DISASSOCIATE IN VIVO. NO PATIENT INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RAD CAP RHEAD RADIAL HEAD AND STEM | PROSTHESIS, ELBOW, HEMI, RADIAL, POLYME | JDB | SMALL BONE INNOVATIONS, INC. | 410-0006/RHA-S4 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |