HEATER/COOLER 3T 120V/60 HZ
Report
- Report Number
- 9611109-2023-00357
- Event Type
- Malfunction
- Date Received
- July 25, 2023
- Report Date
- April 23, 2024
- Manufacturer
- LIVANOVA DEUTSCHLAND GMBH
- Product Code
- DWC
- UDI-DI
- 04033817900528
- PMA / PMN Number
- K191402
- Removal / Correction Number
- Z-0073-2024
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
A.1.-A.5. THERE WAS NO PATIENT INVOLVEMENT. H10: LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN UNITED STATES. LIVANOVA HAS INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
H10: LIVANOVA HAS IDENTIFIED THAT QUICK REFERENCE GUIDE (QRG) VERSION 22 CONTAINS INCONSISTENT INFORMATION REGARDING THE DISINFECTION CALENDAR TABLE IN PAGE 3 OF THE CURRENT USA QRG. THE INFORMATION ON THE QRG IS NOT ALIGNED WITH CURRENT USA VERSION OI OF 3T HEATER COOLER. MEDICAL DEVICE CORRECTION Z-0073-2024 (LIVANOVA REF. FA-CP-MUN-2023-001) WAS DEPLOYED TO INFORM AFFECTED CUSTOMERS, REMOVE ALREADY DISTRIBUTED INCORRECT COPIES AND PROVIDE THE CORRECTED REVISION OF THE QRG. THE RISK IS IN THE ACCEPTABLE REGION. NO SPECIFIC ACTION WAS CURRENTLY DEEMED NECESSARY. LIVANOVA WILL KEEP MONITORING THE MARKET.
LIVANOVA DEUTSCHLAND HAS RECEIVED A REPORT THAT THE CALENDAR SUMMARY TABLE OF THE QUICK REFERENCE GUIDE INCLUDED IN SOME 3T HEATER COOLERS ARE NOT ALIGNED WITH 3T HEATER COOLER IFU. THERE WAS NO PATIENT INVOLVEMENT.
SEE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2193441 | HEATER/COOLER 3T 120V/60 HZ | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND GMBH | 16-02-85 | 04033817900528 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |