FDA Adverse Event Injury Summary report: N

BALL HEADS: COCR BALL HEAD 12/14 Ø 36 SIZE L +3.5

MDR report key: 17387413 · Received July 25, 2023

Report

Report Number
3005180920-2023-00548
Event Type
Injury
Date Received
July 25, 2023
Date of Event
June 27, 2023
Report Date
July 25, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030805226
PMA / PMN Number
K080885
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 4 JULY 2023. LOT 121068: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-APR-2012. EXPIRATION DATE: 2017-FEB-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT (CONSIDERING ALSO RESTERILIZED PIECES) HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDTIONAL DEVICE INVOLVED: BATCH REVIEW PERFORMED ON 4 JULY 2023 ON STEM: AMISTEM H 01.18.132 HA COATED STD STEM SIZE 2 (K093944) LOT. 155651: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 19-JAN-2016. EXPIRATION DATE: 2021-JAN-03. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT (CONSIDERING ALSO RESTERILIZED PIECES) HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

REVISION DUE TO INFLAMMATION FOR SUSPECTED TRUNNIONOSIS, ABOUT 6 YEARS AND 6 MONTHS AFTER THE PRIMARY SURGERY. CUP, LINER, AND HEAD REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632508 BALL HEADS: COCR BALL HEAD 12/14 Ø 36 SIZE L +3.5 HIP COCR BALL HEADS JDI MEDACTA INTERNATIONAL SA 121068 07630030805226

Patients

Seq Age Sex Outcome Treatment
1 71 YR Male Required Intervention