BALL HEADS: COCR BALL HEAD 12/14 Ø 36 SIZE L +3.5
Report
- Report Number
- 3005180920-2023-00548
- Event Type
- Injury
- Date Received
- July 25, 2023
- Date of Event
- June 27, 2023
- Report Date
- July 25, 2023
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JDI
- UDI-DI
- 07630030805226
- PMA / PMN Number
- K080885
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 4 JULY 2023. LOT 121068: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-APR-2012. EXPIRATION DATE: 2017-FEB-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT (CONSIDERING ALSO RESTERILIZED PIECES) HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDTIONAL DEVICE INVOLVED: BATCH REVIEW PERFORMED ON 4 JULY 2023 ON STEM: AMISTEM H 01.18.132 HA COATED STD STEM SIZE 2 (K093944) LOT. 155651: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 19-JAN-2016. EXPIRATION DATE: 2021-JAN-03. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT (CONSIDERING ALSO RESTERILIZED PIECES) HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.
REVISION DUE TO INFLAMMATION FOR SUSPECTED TRUNNIONOSIS, ABOUT 6 YEARS AND 6 MONTHS AFTER THE PRIMARY SURGERY. CUP, LINER, AND HEAD REVISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632508 | BALL HEADS: COCR BALL HEAD 12/14 Ø 36 SIZE L +3.5 | HIP COCR BALL HEADS | JDI | MEDACTA INTERNATIONAL SA | 121068 | 07630030805226 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Male | Required Intervention |