FDA Adverse Event Death Summary report: N

INSYNC SENTRY

MDR report key: 1737628 · Received June 28, 2010

Report

Report Number
6000144-2010-03014
Event Type
Death
Date Received
June 28, 2010
Date of Event
May 6, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S18
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE AND LEADS IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B) (4) BATTERY DEPLETION-NORMAL. (B) (4) NO ANOMALIES FOUND, ALL CONDUCTORS BLOOD/BODY FLUID (NOT OBSTRUCTED), OUTER INSULATION MELTED, APPARENT EXPLANT DAMAGE. FULL LEAD RETURNED AND ANALYZED. (B) (4) PROXIMAL CONDUCTOR FRACTURED, DISTAL CONDUCTOR FRACTURED, DEFIB CONDUCTOR FRACTURED, DISTAL CONDUCTOR BLOOD/BODY FLUID (NOT OBSTRUCTED), DISTAL CONDUCTOR FRACTURE (OVERSTRESS), BLOOD/BODY FLUID OUTER TUBING OVERLAY, INNER TUBING KINKED/BUCKLED, OUTER TUBING OVERLAY MELTED, OUTER TUBING SEPARATION (NON-ELECTRICAL), BLOOD IN/ON HELIX/LOBE MECHANISM, DEFORMATION/FRACTURE IN 5 CM PROXIMAL SECTION OF THE INNER COIL - EXTENSION PROBLEMS. FULL LEAD IN SEGMENTS RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT HAD BEEN REPORTED THE RV (RIGHT VENTRICULAR) LEAD IMPEDANCE HAD INCREASED TO GREATER THAN 1500 OHMS, THERE WAS NOISE, OVERSENSING, FRACTURE, AND INAPPROPRIATE SHOCKS. THE PATIENT'S WIFE LATER REPORTED THAT ON (B)(6) 2010, THE PATIENT WAS TAKEN TO THE HOSPITAL AFTER RECEIVING SEVERAL SHOCKS AND HAD SURGERY TO "REPLACE THE WIRE, BUT DIED TWO DAYS LATER." SHE ALSO REPORTED THE DEATH CERTIFICATE LISTED CAUSE OF DEATH AS "CARDIAC GENETIC SHOCK, SEPTICEMIA OR SOMETHING." THERE IS NO ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT THE DEATH WAS DEVICE RELATED. CAUSE OF DEATH HAS BEEN REQUESTED OF THE PHYSICIAN AND NOT RECEIVED.

Description of Event or Problem · 1

IT HAD BEEN REPORTED THE RV (RIGHT VENTRICULAR) LEAD IMPEDANCE HAD INCREASED TO GREATER THAN 1500 OHMS, THERE WAS NOISE, OVERSENSING, FRACTURE, AND INAPPROPRIATE SHOCKS. THE PATIENT'S WIFE LATER REPORTED THAT ON (B) (6)2010, THE PATIENT WAS TAKEN TO THE HOSPITAL AFTER RECEIVING SEVERAL SHOCKS AND HAD SURGERY TO "REPLACE THE WIRE, BUT DIED TWO DAYS LATER." SHE ALSO REPORTED THE DEATH CERTIFICATE LISTED CAUSE OF DEATH AS "CARDIAC GENETIC SHOCK, SEPTICEMIA OR SOMETHING." THERE IS NO ALLEGATION FROM A HEALTH CARE PROFESSIONAL THAT THE DEATH WAS DEVICE RELATED. CAUSE OF DEATH HAS BEEN REQUESTED OF THE PHYSICIAN AND NOT RECEIVED. FURTHER REPORTED "THE LEAD WAS VERY LIKELY FRACTURED AND PLANS WERE TO REPLACE IT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSYNC SENTRY IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. 7299 ASKU

Patients

Seq Age Sex Outcome Treatment
1 78 YR Death 6935 IMPLANTABLE TACHY LEAD| D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4196 IMPLANTABLE PACING LEAD