FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER

MDR report key: 17374292 · Received July 21, 2023

Report

Report Number
1024879-2023-00488
Event Type
Malfunction
Date Received
July 21, 2023
Date of Event
June 30, 2023
Report Date
July 19, 2023
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903686525
PMA / PMN Number
K991088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: BD RECEIVED 5 SAMPLES FOR INVESTIGATION. THE SAMPLES WERE EVALUATED BY FUNCTIONAL DRAW TESTING AND THE INDICATED FAILURE MODE FOR INSUFFICIENT FLOW WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL PRODUCT SPECIFICATIONS WERE MET. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE INSUFFICIENT BLOOD FLOW. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER CUSTOMER REPORTS LOW DRAW. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED CUSTOMER PROBLEM: CUSTOMER IS REPORTING LOW DRAW VOLUME WITH PRODUCT 368652. LOT 2223194. THEY SEE THE FLASH AND THE BLOOD IS THERE BUT AS SOON AS THEY INSERT THE TUBE IT WOULD BE VERY SLOW TO DRAW AN ONLY FILL UP TO 0.5INCH OF THE TUBE. THEY HAVE TRIED MULTIPLE TUBES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1922522 BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 368652 2223194 50382903686525

Patients

Seq Age Sex Outcome Treatment
1 Unknown