FDA Adverse Event Malfunction Summary report: N

ADMIRAL ACP SYSTEM

MDR report key: 17372603 · Received July 21, 2023

Report

Report Number
3012120772-2023-00013
Event Type
Malfunction
Date Received
July 21, 2023
Date of Event
June 28, 2023
Report Date
July 21, 2023
Manufacturer
SEASPINE ORTHOPEDICS CORPORATION
Product Code
KWQ
UDI-DI
10889981277109
PMA / PMN Number
K212139
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE AP2-100506 TEMPORARY FIXATION PIN AND AP2-100314 ADMIRAL ACP DRILL (REPORT #3012120772-2023-00014) WERE NOT RETURNED FOR ANALYSIS; HOWEVER, REVIEW OF THE X-RAYS PROVIDED CONFIRMS FRAGMENTS OF THE DRILL AND TEMPORARY FIXATION PIN REMAIN IN THE SURGICAL SITE. ADDITIONAL DETAIL REGARDING THE SURGICAL TECHNIQUE AND/OR PATIENT'S ANATOMY HAS NOT BEEN MADE AVAILABLE FOR CONSIDERATION. A DEFINITIVE CAUSE COULD NOT BE ESTABLISHED. REVIEW OF QF-10-01 ADMIRAL ACP SYSTEM IFU: POSSIBLE ADVERSE EVENTS BENDING, DISASSEMBLY, OR FRACTURE OF COMPONENTS.

Description of Event or Problem · 0

ON (B)(6) 2023, A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S ADMIRAL ACP SYSTEM. DURING THE PROCEDURE, AN AP2-100506 TEMPORARY FIXATION PIN WAS INSERTED THROUGH A SCREW HOLE; THE TEMPORARY FIXATION PIN WAS THEN REMOVED WITHOUT BEING AWARE THAT THE PIN HAD FRACTURED AND A FRAGMENT WAS LEFT IN THE SURGICAL SITE. THE SURGEON CONTINUED THROUGH THE PROCEDURE BY INSERTING AN AP2-100314 ADMIRAL ACP DRILL, 14MM INTO THE SAME SCREW HOLE. THE ACP DRILL MADE CONTACT WITH THE TEMPORARY FIXATION PIN AND CREATED ENOUGH FRICTION TO CAUSE IT TO FRACTURE AS WELL. FRAGMENTS OF BOTH THE AP2-100506 TEMPORARY FIXATION PIN AND THE AP2-100314 ACP DRILL REMAIN IN-SITU. NO REPORTS OF TISSUE DAMAGE OR PATIENT INJURY WAS REPORTED AND THERE ARE NO PLANS FOR REVISION OR REMOVAL AT THIS TIME. SEASPINE WAS MADE AWARE OF THIS INCIDENT ON (B)(6) 2023. REPORT # AP2-100506 TEMPORARY FIXATION PIN. REPORT #3012120772-2023-00014 AP2-100314 ACP DRILL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1587327 ADMIRAL ACP SYSTEM TEMPORARY FIXATION PIN KWQ SEASPINE ORTHOPEDICS CORPORATION AP2-100506 10889981277109

Patients

Seq Age Sex Outcome Treatment
1 75 YR Male Other SHORELINE RT| VUMESH