FDA Adverse Event Malfunction Summary report: N

TREO CUSTOM ABDOMINAL STENT-GRAFT SYSTEM

MDR report key: 17371692 · Received July 21, 2023

Report

Report Number
2247858-2023-00187
Event Type
Malfunction
Date Received
July 21, 2023
Date of Event
June 22, 2023
Report Date
December 11, 2023
Manufacturer
BOLTON MEDICAL INC.
Product Code
MIH
PMA / PMN Number
P190015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

BOLTON MEDICAL IS VOLUNTARILY REPORTING AN EVENT RELATED TO A TREO CUSTOM-MADE DEVICE. CUSTOM MADE TREO DEVICES ARE NOT MARKETED IN THE US, HOWEVER THEY ARE SIMILAR TO THE TREO ABDOMINAL STENT-GRAFT SYSTEM APPROVED FOR SALE IN THE US (B)(4). THE EVENT OCCURRED IN THE GERMANY. BOLTON MEDICAL, INC. (D/B/A TERUMO AORTIC), HEREIN KNOWN AS THE "COMPANY", IS SUBMITTING THIS REPORT PURSUANT TO 21 CFR PART 803, HAS MADE REASONABLE EFFORTS TO OBTAIN COMPLETE INFORMATION, AND HAS PROVIDED AS MUCH AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT IS BASED ON INFORMATION OBTAINED BY THE COMPANY, WHICH MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE REGULATIONS. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN, UNAVAILABLE, OR NOT APPLICABLE. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY ANYONE THAT THE DEVICE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE A LEGAL ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY DEFECTS OR HAS MALFUNCTIONED, AS DEFINED FROM A LEGAL STANDPOINT. THESE WORDS ARE INCLUDED IN THE REPORT AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE CERTAIN REGULATIONS, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

BOLTON MEDICAL, INC. (D/B/A TERUMO AORTIC), HEREIN KNOWN AS THE "COMPANY", IS SUBMITTING THIS REPORT PURSUANT TO 21 CFR PART 803, HAS MADE REASONABLE EFFORTS TO OBTAIN COMPLETE INFORMATION, AND HAS PROVIDED AS MUCH AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT IS BASED ON INFORMATION OBTAINED BY THE COMPANY, WHICH MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE REGULATIONS. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN, UNAVAILABLE, OR NOT APPLICABLE. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY ANYONE THAT THE DEVICE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE A LEGAL ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY DEFECTS OR HAS MALFUNCTIONED, AS DEFINED FROM A LEGAL STANDPOINT. THESE WORDS ARE INCLUDED IN THE REPORT AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE CERTAIN REGULATIONS, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

AFTER WITHDRAWING THE BLACK KNOB, THE CAPTURING DID NOT OPEN. SINCE THE GREEN WIRE COULD NOT BE MOVED DISTALLY WITH SMALL PLIERS OR A NEEDLE HOLDER, IT WAS MANIPULATED WITH WIRE PLIERS. THIS IS HOW THE PROBLEM WAS SOLVED. THE PROCEDURE COULD BE COMPLETED WITHOUT FURTHER PROBLEMS. PATIENT OUTCOME - "PATIENT IS DOING WELL."

Description of Event or Problem · 0

AFTER WITHDRAWING THE BLACK KNOB, THE CAPTURING DID NOT OPEN. SINCE THE GREEN WIRE COULD NOT BE MOVED DISTALLY WITH SMALL PLIERS OR A NEEDLE HOLDER, IT WAS MANIPULATED WITH WIRE PLIERS. THIS IS HOW THE PROBLEM WAS SOLVED. THE PROCEDURE COULD BE COMPLETED WITHOUT FURTHER PROBLEMS. PATIENT OUTCOME - "PATIENT IS DOING WELL."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2194114 TREO CUSTOM ABDOMINAL STENT-GRAFT SYSTEM STENT, ENDOVASCULAR GRAFT, AORTIC MIH BOLTON MEDICAL INC. 2304180186

Patients

Seq Age Sex Outcome Treatment
1 74 YR Male Other