H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, ONE HUNDRED (100) RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND TEN (10) WERE SUBJECTED TO FUNCTIONAL TESTING. NO ISSUES WERE OBSERVED RELATING TO UNDERFILL AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT.
H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® K2 EDTA BLOOD COLLECTION TUBES THAT THERE WAS UNDERFILL OR LOW DRAW OF A TUBE WITH BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS REPORTING LOW VACUUM FOR LOT NUMBER 2200102.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® K2 EDTA BLOOD COLLECTION TUBES THAT THERE WAS UNDERFILL OR LOW DRAW OF A TUBE WITH BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS REPORTING LOW VACUUM FOR LOT NUMBER 2200102.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2256702 | BD VACUTAINER® K2 EDTA BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO. (BROKEN BOW) | 367856 | 2200102 | 50382903678568 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |