FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® K2 EDTA BLOOD COLLECTION TUBES

MDR report key: 17368165 · Received July 20, 2023

Report

Report Number
1917413-2023-00676
Event Type
Malfunction
Date Received
July 20, 2023
Date of Event
June 30, 2023
Report Date
July 26, 2023
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903678568
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, ONE HUNDRED (100) RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND TEN (10) WERE SUBJECTED TO FUNCTIONAL TESTING. NO ISSUES WERE OBSERVED RELATING TO UNDERFILL AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® K2 EDTA BLOOD COLLECTION TUBES THAT THERE WAS UNDERFILL OR LOW DRAW OF A TUBE WITH BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS REPORTING LOW VACUUM FOR LOT NUMBER 2200102.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® K2 EDTA BLOOD COLLECTION TUBES THAT THERE WAS UNDERFILL OR LOW DRAW OF A TUBE WITH BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS REPORTING LOW VACUUM FOR LOT NUMBER 2200102.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2256702 BD VACUTAINER® K2 EDTA BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 367856 2200102 50382903678568

Patients

Seq Age Sex Outcome Treatment
1 Unknown