IMP,TSV,MCOL MG,4.7MM,13M
Report
- Report Number
- 0002023141-2023-01964
- Event Type
- Injury
- Date Received
- July 20, 2023
- Date of Event
- June 8, 2023
- Report Date
- January 18, 2024
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019799
- PMA / PMN Number
- K111889
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). A4: WEIGHT UNKNOWN / NOT PROVIDED. G4: ADDITIONAL PMA/510(K) NUMBER ¿ K111889. PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. THE FOLLOWING SECTIONS ARE BEING REPORTED: B4: DATE OF THIS REPORT WAS UPDATED. G3: DATE RECEIVED BY MANUFACTURER WAS UPDATED. G6: TYPE OF REPORT WAS UPDATED. H2: TYPE OF FOLLOW UP WAS UPDATED. H3: DEVICE EVALUATED BY MANUFACTURER WAS UPDATED. H6: TYPE OF INVESTIGATION CODES WERE ADDED: 4109, 4111 AND 3331. H6: INVESTIGATION FINDINGS CODE WAS ADDED: 213. H6: INVESTIGATION CONCLUSIONS CODE WAS ADDED: 67 AND 4315. H10: NARRATIVE/DATA WAS UPDATED. ZIMVIE RECEIVED THE REPORTED IMPLANT FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICE IDENTIFIED THE IMPLANT WITH SIGNS OF USE. IMPLANT WAS RETURNED WITH NO DAMAGE THAT WOULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. MARKINGS LIKELY DUE TO THE REMOVAL PROCESS. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1259399. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1259399 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE MISSING OR CONFUSING INSTRUCTIONS FOR USE, MATERIAL SELECTION. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID NOT OCCUR. THE REPORTED EVENT WAS NON-VERIFIABLE WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
IT WAS REPORTED THAT THE PATIENT CAME TO THE OFFICE WITH INFLAMMATION AND PAIN. CBCT SHOWS RL TO THE BUCCAL AND DISTAL OF THE IMPLANT AT TOOTH LOCATION #30. THE IMPLANT WAS REMOVED. IT APPEARS TO BE A METAL ALLERGY. THE SITE WAS DEBRIDED.
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 910912 | IMP,TSV,MCOL MG,4.7MM,13M | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TSVMWB13 | 1259399 | 00889024019799 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Female | Required Intervention |