FDA Adverse Event Malfunction Summary report: N

RECIPROC BLUE FILES, 6X, STERILE

MDR report key: 17361507 · Received July 20, 2023

Report

Report Number
9611053-2023-00820
Event Type
Malfunction
Date Received
July 20, 2023
Report Date
October 2, 2023
Manufacturer
VDW GMBH
Product Code
EKS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AS A RESULT OF THIS MALFUNCTION, THE POTENTIAL FOR SURGICAL INTERVENTION EXISTS TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION AS EVIDENCED BY PREVIOUS REPORTED EVENTS WITH SIMILAR FILES. THIS EVENT, THEREFORE, IS REPORTABLE PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

SUMMARY: TWO RECIPROC BLUE FILES R25 8/100 25MM WERE RETURNED (ONE FILE IN LOOSE + ONE UNUSED FILE). FILE IN LOOSE IS ACTUALLY BROKEN AT THE TIP OF THE ACTIVE PART (FATIGUE). NO MATERIAL DEFECT WAS FOUND DURING ANALYSIS OF THE RUPTURE PATTERN. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHRS REVIEW (BATCHES #1802037 AND #1803248). UNUSED FILE WAS EVALUATED AND WAS FOUND IN COMPLIANCE WITH SPECIFICATIONS. ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. FOR INFORMATION, WE REMIND THE PRACTITIONER HAS TO MAKE SURE THAT A STRAIGHT-LINE ACCESS IS CREATED PRIOR USING THE RECIPROC BLUE FILES (AS MENTIONED IN THE DFU).

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT RECIPROC BLUE FILES, 6X, STERILE FILE BROKE DURING USE. THE BROKEN PART REMAINS IN THE IN THE ROOT CANAL. FURTHER TREATMENT IS PENDING. FURTHER INFORMATION REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
883605 RECIPROC BLUE FILES, 6X, STERILE FILE, PULP CANAL, ENDODONTIC EKS VDW GMBH 408931

Patients

Seq Age Sex Outcome Treatment
1 Unknown