FDA Adverse Event Death Summary report: N

X SERIES

MDR report key: 17354692 · Received July 19, 2023

Report

Report Number
1220908-2023-02759
Event Type
Death
Date Received
July 19, 2023
Date of Event
June 17, 2023
Report Date
June 28, 2023
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
UDI-DI
00847946019259
PMA / PMN Number
P160022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ZOLL MEDICAL CORPORATION EVALUATED THE DEVICE AND THE CUSTOMER'S REPORT WAS NOT REPLICATED OR CONFIRMED. THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING INCLUDING BENCH HANDLING, SHOCK TESTING, AND ECG STRESS TESTING WITHOUT DUPLICATING THE REPORT. THE DEVICE'S MULTIFUNCTION CABLE (MFC) TO CPR ADAPTOR WERE REPLACED AS A PRE-CAUTION. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. REVIEW OF THE DEVICE LOGS DID NOT REVEAL ANY ASSOCIATED FAULTS WITH THE DEVICE. NO TREND IS ASSOCIATED WITH REPORTS OF THIS TYPE.

Additional Manufacturer Narrative · 0

ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE PRODUCT FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO DEFIBRILLATE A MALE PATIENT (AGE UNKNOWN), THE DEVICE WAS UNABLE TO OBTAIN AN ECG SIGNAL VIA ELECTRODE PADS AND FAILED TO CHARGE. COMPLAINANT INDICATED THAT THE CLINICIAN OBTAINED ANOTHER DEVICE TO CONTINUE TREATING THE PATIENT. COMPLAINANT INDICATED THAT THE PATIENT SUBSEQUENTLY EXPIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35803 X SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION X SERIES NA 00847946019259

Patients

Seq Age Sex Outcome Treatment
1 Male Death