HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Report
- Report Number
- 3005075853-2023-05079
- Event Type
- Malfunction
- Date Received
- July 19, 2023
- Date of Event
- May 23, 2023
- Report Date
- July 19, 2023
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- UDI-DI
- 10705036014621
- PMA / PMN Number
- K132612
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
PC-(B)(4). DATE SENT: 8/3/2023 INVESTIGATION SUMMARY THE PRODUCT WAS RETURNED FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED. THE REMAINING BLADE PORTION WAS SCRATCHED AND WITH EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. DURING FUNCTIONAL TESTING ON GEN11, AN ALERT SCREEN WAS DISPLAYED. PROBABLY THE DEVICE STOPPED ACTIVATING AND DISPLAYED AN ALERT BECAUSE THE KNIFE IS DAMAGED. THE DEVICE WAS DISASSEMBLED TO VERIFY THE INTERNAL COMPONENTS AND NO ANOMALIES WERE FOUND. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES, OR CLIPS DURING THE PROCEDURE, OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE THE SEVERITY OF THE BLADE DAMAGE. THIS IN TURN CAN RESULT IN ACTIVATION ISSUES SUCH AS FAILING THE PRE-RUN TEST WITH THE GENERATOR AND DISPLAYING AN ALERT SCREEN. THESE SCREEN ALERTS CAN BE SUCH AS, ¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE. CONTINUED USAGE OF THE DAMAGED BLADE CAN RESULT IN A BROKEN BLADE. AS PART OF THE QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS.
(B)(4). DATE SENT: 8/28/2023. H6 MEDICAL DEVICE PROBLEM CODE.
(B)(4). DATE SENT: 7/19/2023. D4 BATCH #: X9675A. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT/BATCH. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE ¿RELAX PRESSURE ON BLADE¿ ALERT SCREEN WAS DISPLAYED BY THE GENERATOR AND TITANIUM TIP WAS BROKEN. THE BROKEN PIECE WAS RETRIEVED BY FORCEPS AND WAS DISCARDED. CHANGED TO ANOTHER DEVICE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE TITANIUM TIP BROKE. CHANGED TO ANOTHER DEVICE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 36477 | HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36 | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | HARH36 | X9675A | 10705036014621 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | GENERATOR AND HANDPIECE |