FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION¿ S8 SYSTEM

MDR report key: 17349662 · Received July 19, 2023

Report

Report Number
1723170-2023-01274
Event Type
Malfunction
Date Received
July 19, 2023
Date of Event
July 17, 2023
Report Date
August 18, 2023
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
HAW
UDI-DI
00763000359881
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

H3, H6: A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. TESTING REVEALED THAT REPLACING THE FUSE AND POWER CABLE DIDN¿T RESOLVE THIS ISSUE. THE BATTERY AND UNINTERRUPTIBLE POWER SUPPLY (UPS) ALSO WERE REPLACED. THE SYSTEM THEN PERFORMED AS INTENDED. CODES B01, C02 AND D02 ARE APPLICABLE TO THIS ANALYSIS. THE CABLE, LOT NUMBER: 220424, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE CABLE WAS FOUND TO BE FULLY FUNCTIONAL WITH NO PROBLEM FOUND. THE RETURNED CABLE WAS INTACT AND PASSED A CONTINUITY TEST WITH NO OPENS OR SHORTS DETECTED. THE REPORTED EVENT COULD NOT BE DUPLICATED BY MEDTRONIC PERSONNEL. CODES B01, C19 AND D14 ARE APPLICABLE TO THIS ANALYSIS. THE FUSES WERE RETURNED TO THE MANUFACTURER FOR ANALYSIS. BOTH RETURNED FUSES WERE INTACT. THE FUSES HAVE NOT BLOWN. THE COMPONENT WAS FOUND TO BE FULLY FUNCTIONAL WITH NO PROBLEM FOUND. THE REPORTED EVENT COULD NOT BE DUPLICATED BY MEDTRONIC PERSONNEL. CODES B01, C19 AND D14 ARE APPLICABLE TO THIS ANALYSIS. CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID 9735773 AND 9735787 MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

H2/H6: 9735773, LOT NUMBER#: (B)(4) WAS RETURNED FOR ANALYSIS. THERE WAS NO FAILURE FOUND WITH THE RETURNED PRODUCT. CODES B01, C19, AND D14 APPLY. 9735787, LOT NUMBER#: (B)(4) WAS RETURNED FOR ANALYSIS. AND THE REPORTED ISSUE WAS CONFIRMED. CODES B01, C02, AND D02 APPLY. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM, BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK, BECAUSE THE INFORMATION IS CURRENTLY, UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID 9735881 AND 9735784 H3, H6: A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. TESTING CONFIRMED THAT THE SYSTEM WOULD NOT POWER ON. NO PARTS WERE REPLACED AT THIS TIME. CODES B01, C02 AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE SYSTEM UNEXPECTEDLY SHUT DOWN AND THEN WOULD NOT POWER BACK ON. THE MANUFACTURING REPRESENTATIVE (REP) REPLACED THE FUSES WITH ONES THE BIOMED HAD ON SITE, THE FUSES BLEW IMMEDIATELY. THE REP ALSO PLUGGED IN THE SYSTEM TO A KNOWN WORKING OUTLET AND CHECK THE BACK OF THE UNINTERRUPTIBLE POWER SUPPLY (UPS). THERE WAS NO LEDS ILLUMINATED THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
41902 STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC 9735665 00763000359881

Patients

Seq Age Sex Outcome Treatment
1 Unknown