FDA Adverse Event Malfunction Summary report: N

LP10 VOLUME VENTILATOR

MDR report key: 173478 · Received June 19, 1998

Report

Report Number
2183157-1998-00093
Event Type
Malfunction
Date Received
June 19, 1998
Date of Event
May 20, 1998
Report Date
June 19, 1998
Manufacturer
NELLCOR PURITAN BENNETT
Product Code
CBK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

DURING SERVICE OF PRODUCT FOUND A STOP CYCLE WITH ALL LEDS AND CONSTANT SINGLE TONE ALARM, DUE TO INTEGRATED CIRCUIT U28 ON THE LOGIC BOARD. REPLACED U28.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LP10 VOLUME VENTILATOR VOLUME VENTILATOR CBK NELLCOR PURITAN BENNETT LP10 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other