TALENT THORACIC STENT GRAFT SYSTEM - XCELERANT
Report
- Report Number
- 2953200-2010-01186
- Event Type
- Death
- Date Received
- June 22, 2010
- Date of Event
- May 24, 2010
- Report Date
- May 24, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P070007
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B) (4). EVAL RESULTS: DEATH; ANGULATED AND DISEASED AORTA. CONCLUSION: ANGULATED AND DISEASED AORTA.
A TALENT STENT GRAFT DEVICE WAS IMPLANTED INTO A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 5CM THORACIC AORTIC ANEURYSM. THE AORTA WAS ANGULATED WITH AN 80 DEGREE BEND AND THEN A 60 DEGREE BEND OF THE DESCENDING THORACIC AORTA TOWARD THE CELIAC ARTERY. THE ILIAC ARTERIES WERE SEVERELY TORTUOUS. THE PATIENT HAD A PREVIOUS AORTO-BI-FEMORAL BYPASS FOR AN AAA REPAIR ON 4 MONTHS AGO. FOUR TALENT THORACIC STENT GRAFTS WERE IMPLANTED 1 MONTH AGO USING A CONDUIT IN ORDER TO AVOID INSERTION THROUGH THE AORTO-BI-FEMORAL BYPASS GRAFT. THE PROXIMAL-MOST DEVICE WAS IMPLANTED IN A STRAIGHT AREA OF THE AORTIC ARCH; HOWEVER, DUE TO THE DISTAL AORTIC ANGULATION, THERE WAS DIFFICULTY REMOVING THE REMAINING 3 DELIVERY SYSTEM AFTER THE STENT GRAFTS WERE IMPLANTED AS THE NOSE CONES WERE CATCHING ON THE AORTIC CURVES. FOR THE SECOND DEVICE FROM THE PROXIMAL END THE PHYSICIAN ADVANCED A SHEATH TO CONSTRAIN THE NOSE CONE AND RECAPTURE IT SUCCESSFULLY (SEE MFR # 2953200-2010-01184). THE NOSE CONES OF THE THIRD AND FOURTH DEVICES WERE RECAPTURED BY USING A RELIANT BALLOON TO DEFLECT THE NOSE CONE AWAY FROM THE AORTIC CURVES (SEE MFR #2953200-2010-01185). ALL 3 DEVICES WERE SUCCESSFULLY REMOVED WITH THE NOSE CONES RECAPTURED. THE PATIENT WAS STABLE POST-OPERATIVELY. HOWEVER, THE PATIENT WENT INTO MULTI-ORGAN FAILURE AND EXPIRED THE NEXT DAY POST-OPERATIVELY FROM A MASSIVE ATHEROEMBOLI, WHICH PRESUMABLY CAME DURING THE ENDOVASCULAR REPAIR. THE PATIENT DEVELOPED MULTISYSTEM DYSFUNCTION PRESUMABLY FROM SHOWER EMBOLI DURING ENDOVASCULAR REPAIR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT THORACIC STENT GRAFT SYSTEM - XCELERANT | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00439891 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Death |