BD BBL¿ TRYPTICASE¿ SOY BROTH
Report
- Report Number
- 1119779-2023-00783
- Event Type
- Malfunction
- Date Received
- July 18, 2023
- Date of Event
- July 10, 2023
- Report Date
- September 7, 2023
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- JSG
- UDI-DI
- 30382902218235
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MEDICAL DEVICE EXPIRATION DATE: UNKNOWN A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.
H.6. INVESTIGATION SUMMARY: MATERIAL 221823 IS MANUFACTURED BY REHYDRATING THE MEDIA COMPONENTS WITH USP PURIFIED WATER, AND THOROUGHLY MIXING UNTIL A HOMOGENEOUS SOLUTION IS OBTAINED. THE TUBES ARE TERMINALLY AUTOCLAVED IN AN AIR OVER PRESSURE (AOP) AUTOCLAVE, PER MANUFACTURING INSTRUCTIONS, USING A VALIDATED CYCLE. POST AUTOCLAVING, TUBES ARE PACKAGED INTO FINAL SHIPPING CONFIGURATIONS. THE CUSTOMER DID NOT PROVIDE A BATCH NUMBER, PHOTOS OR RETURNS FOR THIS COMPLAINT INVESTIGATION. TRENDING WAS DONE ON THE APPROPRIATE DATABASES FOR TRYPTICASE SOY BROTH (TSB 15ML) (MATERIAL 221823) AND NO QUALITY NOTIFICATION TRENDS HAVE BEEN IDENTIFIED FOR THIS MATERIAL FOR BROKEN/CRACKED TUBES OR CONTAMINATION IN THE LAST 12 MONTHS. THE COMPLAINT HISTORY WAS REVIEWED FOR MATERIAL 221823 AND THERE ARE NO TRENDS FOR BROKEN OR CRACKED TUBES OR CONTAMINATION IN THE LAST 12 MONTHS. BD WILL CONTINUE TO TREND COMPLAINTS FOR THESE ISSUES. THIS COMPLAINT CANNOT BE CONFIRMED BASED ON MATERIAL TRENDING. NO ADDITIONAL ACTIONS ARE INDICATED AT THIS TIME.
IT WAS REPORTED THAT WHILE USING BD BBL¿ TRYPTICASE¿ SOY BROTH, THERE WAS CONTAMINATION OF AN UNSPECIFIED NUMBER OF PLATES. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTS THAT THEY HAVE PREVIOUS RECEIVED LOTS THAT HAD CONTAMINATED TUBES."
IT WAS REPORTED THAT WHILE USING BD BBL¿ TRYPTICASE¿ SOY BROTH, THERE WAS CONTAMINATION OF AN UNSPECIFIED NUMBER OF PLATES. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTS THAT THEY HAVE PREVIOUS RECEIVED LOTS THAT HAD CONTAMINATED TUBES."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 766548 | BD BBL¿ TRYPTICASE¿ SOY BROTH | CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL | JSG | BECTON DICKINSON & CO. (SPARKS) | 221823 | UNKNOWN | 30382902218235 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |