FS OPT B-KET-NO CODING
Report
- Report Number
- 2954323-2023-31025
- Event Type
- Malfunction
- Date Received
- July 18, 2023
- Date of Event
- June 23, 2023
- Report Date
- July 18, 2023
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- QLG
- PMA / PMN Number
- K193371
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
REPEATED ATTEMPTS BY ADC TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL AND/OR THE CUSTOMER DISCARDED THE PRODUCT. AT THIS TIME PRODUCT HAS NOT YET BEEN RETURNED AND A VALID SERIAL NUMBER HAS NOT BEEN PROVIDED FOR THE LIBRE READER. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. A TRIPPED TREND REVIEW WAS CONDUCTED FOR THE REPORTED COMPLAINT AND LIBRE READER, NO TRENDS WERE IDENTIFIED THAT WOULD INDICATE ANY PRODUCT RELATED ISSUES. THE DHRS (DEVICE HISTORY REVIEW) FOR THE PRECISION STRIPS WAS REVIEWED AND THE DHRS SHOWED THE PRECISION STRIPS PASSED ALL TESTS PRIOR TO RELEASE. RETAIN TESTING WAS PERFORMED FOR PRECISION STRIPS, AND ALL UNITS PERFORMED IN SPECIFICATION AND PASSED. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED, AND A FOLLOW-UP REPORT SUBMITTED. UPON EXTENDED INVESTIGATION, IT WAS DETERMINED THAT THE SERIAL NUMBER PROVIDED BY THE CUSTOMER (B)(6) WAS NOT A VALID SERIAL NUMBER. THEREFORE, SECTION D4 WAS UPDATED TO UNK. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 766499 | FS OPT B-KET-NO CODING | FLASH GLUCOSE MONITORING SYSTEM | QLG | ABBOTT DIABETES CARE INC | 71953-01 | 4500186864 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |