ADVIA CENTAUR XP HBSAG ASSAY
Report
- Report Number
- 1219913-2010-00070
- Event Type
- Other
- Date Received
- June 21, 2010
- Date of Event
- June 2, 2010
- Report Date
- June 7, 2010
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- LOM
- PMA / PMN Number
- P030049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A SIEMENS FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER SITE. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE DISCORDANT (B)(6) RESULT IS UNKNOWN. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
(B)(6) RESULTS WERE OBTAINED FOR TWO PATIENTS SAMPLES. THE LABORATORY'S POLICY IS TO REPEAT ALL HIGH RESULTS. THE PATIENTS SAMPLES WERE REPEATED AND THE RESULTS WERE NEGATIVE. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT (B)(6) RESULTS.
A SIEMENS FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER SITE. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE (B)(6) RESULT IS UNKNOWN. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
(B)(6) RESULTS WERE OBTAINED FOR TWO PATIENTS SAMPLES. THE LABORATORY'S POLICY IS TO REPEAT ALL HIGH RESULTS. THE PATIENTS SAMPLES WERE REPEATED AND THE RESULTS WERE NEGATIVE. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE (B)(6) RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVIA CENTAUR XP HBSAG ASSAY | HBS IMMUNOASSAY | LOM | SIEMENS HEALTHCARE DIAGNOSTICS INC. | NA | 148 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ||||
| 2 |