FDA Adverse Event Malfunction Summary report: N

EMPTY STERILE CONTAINER

MDR report key: 17341724 · Received July 18, 2023

Report

Report Number
1416980-2023-03617
Event Type
Malfunction
Date Received
July 18, 2023
Date of Event
June 20, 2023
Report Date
September 8, 2023
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FKX
UDI-DI
00085412007670
PMA / PMN Number
K833065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E1: INITIAL REPORTER ADDRESS: (B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

D2A: COMMON DEVICE NAME: REMOVE: SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE AND REPLACE WITH SYSTEM, PERITONEAL, AUTOMATIC DELIVERY. D2B: CLASSIFICATION CODE: REMOVE: KDJ AND REPLACE WITH FKX. H10: THE ACTUAL DEVICE WAS NOT AVAILABLE; HOWEVER A PHOTOGRAPH OF THE SAMPLE WAS PROVIDED FOR EVALUATION. THE PHOTOGRAPH WAS VISUALLY INSPECTED AND A SPLIT ON THE PORT OF THE BAG WAS OBSERVED. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE WAS DETERMINED TO BE A MANUFACTURING RELATED ISSUE WHICH OCCURRED DURING THE TUBING ASSEMBLY TO THE BAG. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN EMPTY STERILE CONTAINER LEAKED; FURTHER DESCRIBED AS "LEAKAGE FOUND AT THE CONNECTION SITE" AND "A TEAR WAS SEEN ON CLOSE OBSERVATION". THIS OCCURRED DURING USE OF THE DEVICE FOR PERITONEAL DIALYSIS (PD) THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2201510 EMPTY STERILE CONTAINER SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE CORPORATION NA H23B27044 00085412007670
840714 EMPTY STERILE CONTAINER SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE CORPORATION NA H23B27044 00085412007670

Patients

Seq Age Sex Outcome Treatment
1 Unknown TRANSFER SET