FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 17337431 · Received July 17, 2023

Report

Report Number
1221359-2023-01275
Event Type
Malfunction
Date Received
July 17, 2023
Date of Event
July 10, 2023
Report Date
July 19, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 193606 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 193606, TEST BASE PART NUMBER 195-430H / LOT 183897. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 193606 SHOWED THAT THE COMPLAINT RATE IS (B)(4). IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER, A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.D4 : EXPIRATION DATE H3 OTHER TEXT : SINGLE-USE, DEVICE DISCARDED.

Additional Manufacturer Narrative · 0

THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION OR UPON RECEIPT OF NEW INFORMATION. H3 OTHER TEXT : SINGLE-USE, DEVICE DISCARDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 AG SELF TEST PERFORMED ON (B)(6) 2023 ON A NASAL SAMPLE. REPEAT TESTING WAS PERFORMED ON (B)(6) 2023 AND GENERATED A NEGATIVE RESULT. ADDITIONAL TESTING WAS PERFORMED ON (B)(6) 2023 WITH A NON-ABBOTT RAPID TEST IN A FACILITY AND GENERATED A NEGATIVE RESULT. THE CONSUMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Description of Event or Problem · 0

THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 AG SELF TEST PERFORMED ON (B)(6) 2023 ON A NASAL SAMPLE. REPEAT TESTING WAS PERFORMED ON (B)(6) 2023 AND GENERATED A NEGATIVE RESULT. ADDITIONAL TESTING WAS PERFORMED ON (B)(6) 2023 WITH A NON-ABBOTT RAPID TEST IN A FACILITY AND GENERATED A NEGATIVE RESULT. THE CONSUMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1628116 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 193606 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 77 YR Female