FDA Adverse Event Malfunction Summary report: N

TOTALCARE-TREAT SURFACE&SCALE TOTAL CARE FRAME

MDR report key: 1733164 · Received June 15, 2010

Report

Report Number
1824206-2010-08236
Event Type
Malfunction
Date Received
June 15, 2010
Date of Event
January 27, 2008
Report Date
January 27, 2008
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN REPLACED THE LEFT SIDERAIL PATIENT CONTROL BOARD TO RESOLVE THE ISSUE. PURSUANT TO OUR RESPONSE TO WARNING LETTER 2008-DT-04, HILL-ROM IS CONTINUING ITS RETROSPECTIVE REVIEW OF EVENTS FOR REPORTABILITY. THIS EFFORT WILL CONTINUE UNTIL WE ARE CURRENT.

Description of Event or Problem · 1

TECHNICIAN ALLEGED THAT THE HEAD SECTION WAS GOING UP BY ITSELF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE-TREAT SURFACE&SCALE TOTAL CARE FRAME A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. P1900AF1619

Patients

Seq Age Sex Outcome Treatment
1