FDA Adverse Event Injury Summary report: N

LO-PRO SCRW TM,SS 3.5X 16MMCORT

MDR report key: 17326428 · Received July 14, 2023

Report

Report Number
1220246-2023-07164
Event Type
Injury
Date Received
July 14, 2023
Date of Event
May 2, 2023
Report Date
October 27, 2023
Manufacturer
ARTHREX, INC.
Product Code
HWC
UDI-DI
00888867047297
PMA / PMN Number
K203239
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT ALLEGATION WAS CONFIRMED. ONLY WAS SCREW WAS RECEIVED FOR THIS SIZE. ONE UNPACKAGED AR-8835-16 LOW PROFILE SCREW¿, SS, 3.5 X 16 MM, CORTICAL, BATCH UNK, WAS RETURNED FOR INVESTIGATION. THE RETURNED SCREW WAS VISUALLY INSPECTED, AND IT WAS NOTED NO VISUAL DEFECTS. UPON FUNCTIONAL TESTING WITH THE RETURNED MATING PART AR-2652CL, THE LOW-PROFILE SCREW FIT LOOSE ON THE LOCKING HOLES. HOWEVER, THE SCREW HAS THE CORRECT SIZE PER DRAWING C18816 AT REVISION 1. THE METHOD OF BONE PREPARATION EMPLOYED DURING THE PROCEDURE, AS WELL AS THE BONE QUALITY ENCOUNTERED, WAS NOT PROVIDED. THE MOST LIKELY CAUSE FOR THE REPORTED FAILURE IS A USER ERROR. PER DFU-0125 AT REV. 2. G. 1. SURGEONS MUST APPLY THEIR PROFESSIONAL JUDGMENT WHEN DETERMINING THE APPROPRIATE SCREW SIZE BASED ON THE SPECIFIC INDICATION, PREFERRED SURGICAL TECHNIQUE, AND PATIENT HISTORY. 2. SURGEONS ARE ADVISED TO REVIEW THE PRODUCT-SPECIFIC SURGICAL TECHNIQUE PRIOR TO PERFORMING ANY SURGERY. ARTHREX PROVIDES DETAILED SURGICAL TECHNIQUES IN PRINT, VIDEO, AND ELECTRONIC FORMATS. THE ARTHREX WEBSITE ALSO PROVIDES DETAILED SURGICAL TECHNIQUE INFORMATION AND DEMONSTRATIONS. OR CONTACT YOUR ARTHREX REPRESENTATIVE FOR AN ONSITE DEMONSTRATION. 3. USE THE APPROPRIATELY SIZED DRILL BIT FOR THE SCREW. 4. DAMAGE TO THE DRIVER OR SCREW MAY RESULT FROM FAILURE TO SEAT THE DRIVER FULLY INTO THE SCREW OR TO ALIGN THE DRIVER PROPERLY WITH THE SCREW.

Additional Manufacturer Narrative · 0

THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT HAS NOT YET BEEN DETERMINED AS THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED, INCLUDING A MOST LIKELY CAUSE IF A ROOT CAUSE CAN NOT BE DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SCREW HAS LOOSENED IN THE CLAVICLE PLATE. NO FURTHER INFORMATION RECEIVED. 27-JUN-2023 UPDATE DW: FURTHER INFORMATION WERE PROVIDED THAT THE FOLLOWING DEVICES BECAME LOOSE POST OP. AR-2652CL BATCH: 5262102; AR-8835L-14 BATCH: 15040867; AR-8835-14 BATCH: 10694746; 2X AR-8835-16 BATCH: 1072011; 3X AR-8835-18 BATCH: 10591283; AR-8827-14 BATCH: 1074240. AS THE REPORTED DEVICES ARE AVAILABLE FOR RETURN A REVISION SURGERY HAD TO BE PERFORMED BUT NO FURTHER INFORMATION WERE PROVIDED WHEN THE REVISION SURGERY WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
805453 LO-PRO SCRW TM,SS 3.5X 16MMCORT BONE FIXATION SCREW/PLATE HWC ARTHREX, INC. LO-PRO SCRW TM,SS 3.5X 16MMCORT UNK 00888867047297

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other