FDA Adverse Event Malfunction Summary report: N

MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 17323713 · Received July 14, 2023

Report

Report Number
2032227-2023-237011
Event Type
Malfunction
Date Received
July 14, 2023
Date of Event
June 25, 2023
Report Date
August 29, 2023
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000414344
PMA / PMN Number
P160017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE PUMP PASSED THE DISPLACEMENT TEST AND ACTIVE CURRENT TEST. THE PUMP FAILED SLEEP CURRENT TEST. THE PUMP FAILED THE SELF TEST DUE TO PUMP ERROR 75. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUMP. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE ON PCBA 1, PCBA 2, MOTOR, LCD CONNECTOR, AND LITHIUM BATTERY CONNECTOR. THE FOLLOWING POWER ALARMS/ALERTS NOTED IN DOWNLOADED HISTORY ON EVENT DATE: REPLACE BATTERY ALERT [73] ON (B)(6) 2023 18:15:46.000, REPLACE BATTERY NOW ALARM [11] ON (B)(6) 2023 18:04:54.000, AND BATTERY FAILED ALARM [58] ON (B)(6) 2023 18:04:53.000. THE POWER MANAGEMENT GRAPH CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. THE FOLLOWING ALARMS/ALERTS WERE NOTED ON EVENT DATE: PUMP ERROR 63 ON (B)(6) 2023 18:04:50.000, PUMP ERROR 4 ON (B)(6) 2023 18:27:15.000, AND PUMP ERROR 75 ON (B)(6) 2023 07:53:22.000. CONFIRMED PUMP ERROR 63 (FILENUMBER = 2101 LINENUMBER = 249) DUE TO HARDWARE ERROR. PUMP ERROR 4 WAS A CONSEQUENCE OF PUMP ERROR 63. PUMP ERROR 75 CONFIRMED DUE TO MOISTURE DAMAGE. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: BATTERY TUBE THREADS - CRACKED, CRACKED KEYPAD OVERLAY, STAINED KEYPAD OVERLAY, AND PILLOWING KEYPAD OVERLAY. NO BLANK DISPLAY ANOMALIES NOTED DURING ANALYSIS, BLANK DISPLAY NOT CONFIRMED. CONFIRMED PUMP ERROR 63 DUE TO HARDWARE ERROR, ISOLATED TO ELECTRONIC STACK. PUMP ERROR 4 WAS A CONSEQUENCE OF PUMP ERROR 63. PUMP ERROR 75 CONFIRMED DUE TO MOISTURE DAMAGE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THAT THE PUMP QUIT WORKING AND DOESN'T TURN ON. TROUBLESHOOTING WAS PERFORMED AND IT WAS INDICATED THAT THE DISPLAY WAS BLANK. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE DEVICE. THE PRODUCT WILL BE RETURNED FOR FAILURE ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
699478 MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1880 HG54BBQZZ 000000763000414344

Patients

Seq Age Sex Outcome Treatment
1 62 YR Female