FDA Adverse Event Malfunction Summary report: N

BUBBLE DETECTOR SENSOR

MDR report key: 17323000 · Received July 14, 2023

Report

Report Number
9611109-2023-00330
Event Type
Malfunction
Date Received
July 14, 2023
Date of Event
June 14, 2023
Report Date
July 13, 2023
Manufacturer
LIVANOVA DEUTSCHLAND
Product Code
KRL
UDI-DI
04033817900573
PMA / PMN Number
K955152
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H10: THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT THE SENSOR REACTED EVEN THOUGH THERE WAS NO AIR BUBBLES. IN CASE OF BUBBLE SENSOR TRIGGERING FALSE BUBBLE ALARMS (OVER-SENSING ISSUE), THE PUMP IS STOPPED BY THE BUBBLE ALARM EVEN IF NO AIR IS PRESENT IN THE MONITORED LINE. IN SUCH SITUATIONS, THE ALARM CAN ALWAYS BE CLEARED/OVERRIDDEN BY USER, ONCE THE LINE IS VERIFIED TO BE FREE OF AIR, AND THE PERFUSION CAN BE EASILY RESTARTED. THUS, IT IS UNLIKELY THAT A FALSE BUBBLE ALARM WILL RESULT IN SERIOUS INJURY. BASED ON THE ABOVE RATIONALE, THE EVENT HAS BEEN RE-ASSESSED AS NOT REPORTABLE.

Additional Manufacturer Narrative · 0

THERE WAS NO PATIENT INVOLVEMENT. LIVANOVA DEUTSCHLAND MANUFACTURES THE BUBBLE DETECTOR SENSOR. THE INCIDENT OCCURRED IN KURUME, JAPAN. THROUGH FOLLOW-UP COMMUNICATION WITH LIVANOVA FIELD SERVICE REPRESENTATIVE, A NEW BUBBLE DETECTOR SENSOR WILL BE SHIPPED TO THE CUSTOMER. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

SEE INITIAL REPORT.

Description of Event or Problem · 0

LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A BUBBLE SENSOR DETECTOR MALFUNCTIONED DURING PRIMING. NO ADDITIONAL INFORMATION IS AVAILABLE. IT CANNOT BE EXCLUDED THAT BUBBLE SENSOR DID NOT DETECT AIR BUBBLE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
699435 BUBBLE DETECTOR SENSOR MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS KRL LIVANOVA DEUTSCHLAND 23-07-40 04033817900573

Patients

Seq Age Sex Outcome Treatment
1 Unknown