UNIPOLAR BOVIE CORD W/UNIVERSAL PLUG
Report
- Report Number
- 1423507-2023-00094
- Event Type
- Malfunction
- Date Received
- July 13, 2023
- Date of Event
- June 22, 2023
- Report Date
- August 16, 2023
- Manufacturer
- CAREFUSION, INC
- Product Code
- GEI
- UDI-DI
- 10885403154805
- PMA / PMN Number
- K971540
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
(B)(4). FOLLOW UP THE COMPLAINT PRODUCT WAS RETURNED, AND AN EVALUATION WAS PERFORMED. THE RETURNED DEVICE WAS REVIEWED AND WAS CONFIRMED TO HAVE SIGNS OF SIGNIFICANT WEAR AND TEAR. THE LOT NUMBER WAS CONFIRMED TO BE 100435 WHICH WAS SOLD TO CAREFUSION/BD IN 2017. THIS DEVICE IS ONLY VALIDATED FOR 20 STERILIZATIONS/USES. THE PRODUCT AT THE TIME OF MANUFACTURING WOULD HAVE CONFORMED TO ALL SPECIFICATIONS FOR THE PRODUCTION. THE ROOT CAUSE FOR THE REPORTED ISSUE IS MOST LIKELY DUE TO WEAR DURING MAINTENANCE/PROCESSING, EXCESSIVE USE AND STERILIZATION, AND APPEARS TO HAVE REACHED END OF LIFE. THERE HAVE BEEN NO ISSUES IDENTIFIED WITH THE MATERIAL OR MANUFACTURING PROCESS.
(B)(4). 28JUN2023 WRITER SENT THE CUSTOMER AN EMAIL ACKNOWLEDGING RECEIPT OF THE COMPLAINT AND REQUESTED FOLLOW UP INFORMATION INCLUDING AS TO IF THERE WAS ANY PATIENT IMPACT RELATED TO THIS EVENT. WRITER PROVIDED CONTACT INFORMATION. DEVICE NOT YET EVALUATED, IF THE DEVICE IS EVALUATED A FOLLOW UP WILL BE SENT.
VERBATIM: CUSTOMER REPORTED THE CORD STARTED TO CHANGE TO A GLOWING ORANGE COLOR. THE END THAT CONNECTS INTO THE INSTRUMENT THAT LEADS TO THE FLEXIBLE PORTION OF THE HUB ALSO NEAR GRAY CORD WHERE CORD STOPS AT BLUEISH CONNECTER STARTED TO GLOW ORANGE. THE CUSTOMER REPORTS THAT THEY LEFT CORD IN BIOHAZARD BAG. (B)(6) 2023 ¿ WHAT IS THE MATERIAL NUMBER? 88-9199 ¿ WHAT IS THE LOT NUMBER? 100435 ¿ WAS THE PRODUCT RECEIVED IN THIS CONDITION? NO ¿ PLEASE CONFIRM WHETHER OR NOT THERE WAS ANY PATIENT IMPACT. NO HARM OR INJURY TO PATIENT ¿ DO YOU HAVE PHOTOS SHOWING THE REPORTED ISSUE? YES ¿ ATTACHED TO THE EMAIL ¿ IS THE PRODUCT AVAILABLE TO SEND BACK TO US FOR AN EVALUATION? YES ¿ PLEASE PROVIDE DATE OF INCIDENT. (B)(6) 2023. ADDITIONAL INFORMATION: YES THE CORD CHANGED TO A GLOWING ORANGE COLOR BECAUSE IT GOT TOO HOT. THE CORD STARTED TO GLOW ORANGE IN COLOR, BUT ONCE THE CORD SEPARATED, IT WENT BACK TO THE ORIGINAL GREY COLOR. NO FURTHER INFORMATION IS NOT AVAILABLE.
VERBATIM: CUSTOMER REPORTED THE CORD STARTED TO CHANGE TO A GLOWING ORANGE COLOR. THE END THAT CONNECTS INTO THE INSTRUMENT THAT LEADS TO THE FLEXIBLE PORTION OF THE HUB ALSO NEAR GRAY CORD WHERE CORD STOPS AT BLUEISH CONNECTER STARTED TO GLOW ORANGE. THE CUSTOMER REPORTS THAT THEY LEFT CORD IN BIOHAZARD BAG. WED 06/28/2023 ¿ WHAT IS THE MATERIAL NUMBER? 88-9199 ¿ WHAT IS THE LOT NUMBER? 100435 ¿ WAS THE PRODUCT RECEIVED IN THIS CONDITION? NO ¿ PLEASE CONFIRM WHETHER OR NOT THERE WAS ANY PATIENT IMPACT. NO HARM OR INJURY TO PATIENT ¿ DO YOU HAVE PHOTOS SHOWING THE REPORTED ISSUE? YES ¿ ATTACHED TO THE EMAIL. ¿ IS THE PRODUCT AVAILABLE TO SEND BACK TO US FOR AN EVALUATION? YES. ¿ PLEASE PROVIDE DATE OF INCIDENT. (B)(6) 2023. 07JUL2023 ADDITIONAL INFORMATION: YES THE CORD CHANGED TO A GLOWING ORANGE COLOR BECAUSE IT GOT TOO HOT. THE CORD STARTED TO GLOW ORANGE IN COLOR, BUT ONCE THE CORD SEPARATED, IT WENT BACK TO THE ORIGINAL GREY COLOR. NO FURTHER INFORMATION IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 840791 | UNIPOLAR BOVIE CORD W/UNIVERSAL PLUG | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | CAREFUSION, INC | 88-9199 | 100435 | 10885403154805 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |