FDA Adverse Event Malfunction Summary report: N

SUPERDIMENSION INREACH BRONCHUS SYSTEM (SDBS)

MDR report key: 1731862 · Received June 10, 2010

Report

Report Number
3004962788-2010-00021
Event Type
Malfunction
Date Received
June 10, 2010
Date of Event
May 14, 2010
Report Date
June 10, 2010
Manufacturer
SUPERDIMENSION INC.
Product Code
JAK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVAL: A SERVICE CALL WAS PERFORMED AT THIS SITE ON (B)(4), 2010. A COMPONENT OF THE SYSTEM, THE UNION, WAS REPLACED. THE UNION WAS RETURNED FOR ANALYSIS AND THE REPORT SHOWED THAT THE LOCATION BOARD CONNECTOR ON THE UNION WAS CRACKED WHICH LED TO INTERMITTENT ELECTRICAL CONNECTION. THERE WAS NO INJURY TO THE PATIENT REPORTED. IN AN ABUNDANCE OF CAUTION, THIS EVENT IS BEING REPORTED DUE TO THE ADD'L POTENTIAL RISK ASSOCIATED WITH MULTIPLE EXPOSURES TO GENERAL ANESTHESIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LOCATION BOARD, A COMPONENT OF THE SUPERDIMENSION SYSTEM, WOULD NOT CONNECT TO THE SYSTEM. THE SUPERDIMENSION PORTION OF THE CASE WAS CANCELLED WITH THE PT UNDER GENERAL ANESTHESIA. THERE WAS NO HARM OR INJURY TO THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUPERDIMENSION INREACH BRONCHUS SYSTEM (SDBS) COMPUTED TOMOGRAPHY X-RAY SYSTEM JAK SUPERDIMENSION INC. I5102045

Patients

Seq Age Sex Outcome Treatment
1