FDA Adverse Event Malfunction Summary report: N

SPECTRUM IQ INFUSION PUMP

MDR report key: 17313785 · Received July 13, 2023

Report

Report Number
1314492-2023-02516
Event Type
Malfunction
Date Received
July 13, 2023
Report Date
September 27, 2023
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K222048
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION D9, H3, H6 AND H10: THE DEVICE WAS RECEIVED FOR EVALUATION. DURING FUNCTIONAL TESTING, DISPLAY IS CUTTING OUT WAS REPRODUCED AS THE LIQUID CRYSTAL DISPLAY (LCD) WAS TOO BRIGHT AND DEFECTIVE. A SERVICE HISTORY REVIEW WAS PERFORMED AND REVEALED THAT THE DEVICE HAS NO PREVIOUS SERVICE EVENTS; THEREFORE, SERVICING DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE A FAILED LCD. THE LCD REQUIRES REPLACEMENT TO ADDRESS THIS ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE DISPLAY OF SPECTRUM IQ PUMP WAS CUTTING OUT. THIS OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2209979 SPECTRUM IQ INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION NA NA

Patients

Seq Age Sex Outcome Treatment
1 Unknown