TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2010-01163
- Event Type
- Death
- Date Received
- June 18, 2010
- Date of Event
- May 19, 2010
- Report Date
- May 19, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B) (4). EVAL RESULTS: DEATH, ENDOLEAK, ARTERIAL AND VENOUS OCCLUSION; SHORT AORTIC NECK; DEVICE WAS USED IN A PATIENT WITH A INADEQUATE VESSEL MORPHOLOGY.
A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 7 CM DIAMETER ABDOMINAL AORTIC ANEURYSM. THE AORTIC NECK WAS 7 MM LONG AND 41.8 MM IN DIAMETER AT 10 MM BELOW THE RENALS. THE NECK ALSO HAD MORE THAN 50% THROMBUS AROUND THE RENALS. THE PATIENT HAD CO-MORBIDITIES, AND WAS NOT AN OPEN REPAIR CANDIDATE. THE VESSELS HAD SOME TORTUOSITY AND CALCIFICATION. THE PHYSICIAN COMMENTED THAT A TYPE 1 ENDOLEAK WOULD LIKELY OCCUR POST IMPLANT DUE TO THE PATIENT'S VESSEL MORPHOLOGY. THE TALENT BIFURCATED STENT GRAFT WAS IMPLANTED VERY PROXIMALLY DUE TO THE SHORT NECK AND LANDED INTENTIONALLY RIGHT BELOW THE SMA, SO THE RENALS WERE COVERED. A RENAL-FEMORAL BYPASS WAS THEN PERFORMED ON THE LEFT SIDE TO PERFUSE THE LEFT RENAL ARTERY; HOWEVER, THE PATIENT MAY HAVE BEEN WITHOUT RENAL FLOW FOR UP TO 45 MINUTES. A PROXIMAL TYPE 1 ENDOLEAK WAS PRESENT AT THE END OF THE PROCEDURE, BUT THERE WAS NO ROOM PROXIMALLY TO PLACE A CUFF OR PALMAZ. THE ENDOLEAK WAS LEFT UNTREATED, BUT THE PATIENT WAS STABLE. THE PATIENT REPORTEDLY EXPIRED 3 DAYS POST-IMPLANT FROM CO-MORBIDITIES. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00451816 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 88 YR | Death |