FDA Adverse Event Injury Summary report: N

ICHEM® VELOCITY¿ AUTOMATED URINE CHEMISTRY SYSTEM

MDR report key: 17307570 · Received July 12, 2023

Report

Report Number
2122870-2023-00083
Event Type
Injury
Date Received
July 12, 2023
Date of Event
June 14, 2023
Report Date
July 12, 2023
Manufacturer
BECKMAN COULTER IRELAND INC.
Product Code
KQO
UDI-DI
10837461001751
PMA / PMN Number
K171083
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BEC FIELD SERVICE ENGINEER (FSE) WAS AT THE CUSTOMER SITE AND OBSERVED THAT THE ICHEM VELOCITY INSTRUMENT WAS NOT DOSING THE VELOCITY URNALYSIS STRIPS PROPERLY. THE FSE FOUND THAT THE SYRINGE PUMP WAS FAULTY, THE RUBBER STOPPER WAS STUCK INSIDE THE SYRINGE AND WOULD NOT ALLOW THE INSTRUMENT TO FULLY DOSE THE PADS. TO RESOLVE THE ISSUE THE FSE REPLACED THE SYRINGE PUMP, CHECKED FLUID CONNECTIONS AND SPM ALIGNMENTS. TO VERIFY REPAIRS, CONTROLS WERE RUN AND PASSED SUCCESSFULLY. BEC INTERNAL IDENTIFIER - (B)(4). RELATED EVENTS: (B)(4).

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSE NEGATIVE ERRONEOUS RESULTS FOR URINE SAMPLES RUN ON THEIR ICHEM VELOCITY URINE CHEMISTRY ANALYZER. NO FLAGS WERE GENERATED BY INSTRUMENT. THIRTEEN (13) PATIENT RESULTS WERE CONSIDERED TO BE IN QUESTION AND THE CUSTOMER BEGAN URINE CULTURES FOR THEM. THE ORIGINAL SAMPLES THAT HAD BEEN REFRIGERATED WERE RE-RUN WITHIN 24 HOURS. THE CULTURE RESULTS FOR SAMPLE 122316401328 SHOWED 50,00-100,000 STAPHYLOCOCCUS AND THE PATIENT WAS PRESCRIBED ANTIBIOTICS WITHIN 24 HOURS FROM THE TIME THE ORIGINAL URINALYSIS COLLECTION TOOK PLACE. THERE WERE ANOTHER FIVE (5) PATIENTS THAT REQUIRED ANTIBIOTICS, ONE OF WHICH WAS PRESCRIBED ANTIBIOTICS FOR BACTEREMIA/SEPSIS PRIOR TO THE URINE CULTURE BEING PROCESSED. INDIVIDUAL REPORTS WILL BE FILED FOR THE OTHER FOUR (4) PATIENTS AND WILL BE REFERENCED AS RELATED EVENTS IN SECTION H10. THERE WAS NO ADVERSE EVENT REPORTED FOR ANY OF THE 5 PATIENTS THAT WERE DELAYED IN HAVING ANTIBIOTICS PRESCRIBED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
190513 ICHEM® VELOCITY¿ AUTOMATED URINE CHEMISTRY SYSTEM AUTOMATED URINALYSIS SYSTEM KQO BECKMAN COULTER IRELAND INC. 2.5 ICHEM VELOCITY INSTRUMENT COMPUTERLESS BRING UP TESTED I 10837461001751

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other