FDA Adverse Event Malfunction Summary report: N

BD BBL CHOCOLATE II AGAR, 100 PLATES

MDR report key: 17299033 · Received July 11, 2023

Report

Report Number
1119779-2023-00760
Event Type
Malfunction
Date Received
July 11, 2023
Date of Event
June 20, 2023
Report Date
July 13, 2023
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JTY
PMA / PMN Number
PREAMENDMENT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS HAVE BEEN UPDATED WITH CORRECTED AND/OR ADDITIONAL INFORMATION. D9: DEVICE AVAILABLE FOR EVALUATION: YES, D9: RETURNED TO MANUFACTURER ON: 2023-07-11. H.6. INVESTIGATION SUMMARY: MATERIAL #: 251267; LOT/BATCH #: 3101725. BD RECEIVED CUSTOMER RETURNED SAMPLES AND PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR BIOLOGICAL CONTAMINATION WITH THE INCIDENT LOT WAS NOT OBSERVED. ADDITIONALLY, THE CUSTOMER SAMPLES ALONG WITH RETENTION SAMPLES FROM BD INVENTORY, WERE EVALUATED AND NO ISSUES WERE IDENTIFIED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR CONTAMINATION. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BBL CHOCOLATE II AGAR, 100 PLATES THERE WAS CONTAMINATION FOUND ON THE MEDIA. THIRTY PLATES WERE AFFECTED. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ACCORDING TO THE CUSTOMER'S REPORT, SIGNS OF WHITISH MATTER WERE FOUND ON THE SURFACE OF THE MEDIA."

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BBL CHOCOLATE II AGAR, 100 PLATES THERE WAS CONTAMINATION FOUND ON THE MEDIA. THIRTY PLATES WERE AFFECTED. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ACCORDING TO THE CUSTOMER'S REPORT, SIGNS OF WHITISH MATTER WERE FOUND ON THE SURFACE OF THE MEDIA."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1569391 BD BBL CHOCOLATE II AGAR, 100 PLATES CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA JTY BECTON DICKINSON & CO. (SPARKS) 3101725

Patients

Seq Age Sex Outcome Treatment
1 Unknown