FDA Adverse Event Injury Summary report: N

HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER

MDR report key: 1729714 · Received June 16, 2010

Report

Report Number
1226348-2010-00187
Event Type
Injury
Date Received
June 16, 2010
Manufacturer
CODMAN & SHURTLEFF, INC., MEDOS S.A.
Product Code
JXG
PMA / PMN Number
K974739
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON COMPLETION OF THE INVESTIGATION, IT WAS NOTED THAT THIS COMPLAINT COULD NOT BE CONFIRMED. THE DEVICE WAS TESTED AND PASSED ALL TESTS. THE PROBLEM REPORTED BY THE CUSTOMER COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. A REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THAT THE DEVICE CONFORMED TO THE REQUIRED SPECIFICATIONS PRIOR TO DISTRIBUTION. BASED ON THE RESULTS OF THIS INVESTIGATION, NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME, THIS COMPLAINT IS CLOSED.

Description of Event or Problem · 1

AFFILIATE REPORTED THAT FLUID DID NOT FLOW THROUGH THE DEVICE AFTER IMPLANTATION AND AS A RESULT, THE DEVICE WAS REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HAKIM PROGRAMMABLE VALVE WITH PRECHAMBER SHUNT, CENTRAL NERVOUS SYSTEM & COMPS JXG CODMAN & SHURTLEFF, INC., MEDOS S.A. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention