FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 17295868 · Received July 11, 2023

Report

Report Number
9610877-2023-56301
Event Type
Malfunction
Date Received
July 11, 2023
Date of Event
June 29, 2023
Report Date
July 11, 2023
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FAJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL FB-15V IS AVAILABLE IN THE USA WITH A 510K NUMBER K951199. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OCULAR FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE OCULAR. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
312321 PENTAX FIBEROPTICCYSTOSCOPE 2.1C 4.9TP 400L ROP FAJ HOYA CORPORATION PENTAX TOKYO OFFICE FCY-15RBS

Patients

Seq Age Sex Outcome Treatment
1 Unknown