BD INSULIN SYRINGES 3/10ML 6MM (15/64") 31G U-100
Report
- Report Number
- 1920898-2023-00432
- Event Type
- Malfunction
- Date Received
- July 10, 2023
- Date of Event
- June 19, 2023
- Report Date
- October 5, 2023
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE FOLLOWING FIELDS HAVE BEEN UPDATED DUE TO ADDITIONAL INFORMATION: D.9. DEVICE AVAILABLE FOR EVAL?: YES D.9. RETURNED TO MANUFACTURER ON: 14-JUL-2023 H.6. INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AND BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED.
B.3. DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: D.4. MEDICAL DEVICE LOT #: 1109192, D.4. MEDICAL DEVICE EXPIRATION DATE: 30-APR-2026, H.4. DEVICE MANUFACTURE DATE: 19-APR-2021. D.4. MEDICAL DEVICE LOT #: 1326350, D.4. MEDICAL DEVICE EXPIRATION DATE: 31-DEC-2026, H.4. DEVICE MANUFACTURE DATE: 22-NOV-2021. D.4. MEDICAL DEVICE LOT #: 2066145, D.4. MEDICAL DEVICE EXPIRATION DATE: 31-MAR-2027, H.4. DEVICE MANUFACTURE DATE: 07-MAR-2022. D.4. MEDICAL DEVICE LOT #: 2220410, D.4. MEDICAL DEVICE EXPIRATION DATE: 31-AUG-2027, H.4. DEVICE MANUFACTURE DATE: 08-AUG-2022. H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT DURING USE WITH AN UNSPECIFIED AMOUNT OF BD INSULIN SYRINGES 3/10ML 6MM THE SYRINGE IS DAMAGED AND APPEARS BENT RESULTING IN COMPONENT SEPARATION. 2 OF 2 REPORTED DEFECTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE PURCHASED 324919 SYRINGES FROM MULTIPLE LOTS AND APPROXIMATELY SINCE APRIL 2022 WE STARTED SEEING DEFECTIVE SYRINGES IN EACH LOT WE RECEIVED. ... WE SAW SYRINGE HEADS FALL OFF DURING NORMAL USE IN EXTREME CASES OF THE "HEAD GAP" DEFECT.
IT WAS REPORTED THAT DURING USE WITH AN UNSPECIFIED AMOUNT OF BD INSULIN SYRINGES 3/10ML 6MM THE SYRINGE IS DAMAGED AND APPEARS BENT RESULTING IN COMPONENT SEPARATION. 2 OF 2 REPORTED DEFECTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE PURCHASED 324919 SYRINGES FROM MULTIPLE LOTS AND APPROXIMATELY SINCE APRIL 2022 WE STARTED SEEING DEFECTIVE SYRINGES IN EACH LOT WE RECEIVED. ... WE SAW SYRINGE HEADS FALL OFF DURING NORMAL USE IN EXTREME CASES OF THE "HEAD GAP" DEFECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 312283 | BD INSULIN SYRINGES 3/10ML 6MM (15/64") 31G U-100 | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |