FDA Adverse Event Malfunction Summary report: N

BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM

MDR report key: 17293095 · Received July 10, 2023

Report

Report Number
3014704491-2023-00333
Event Type
Malfunction
Date Received
July 10, 2023
Date of Event
June 14, 2023
Report Date
July 30, 2023
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 3080105. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. BASED ON THE DESCRIPTION OF THE EVENT, OUR ENGINEERS WERE ABLE TO DETERMINE THAT THE MOST LIKELY ROOT CAUSE FOR THIS EVENT IS THE FORCING OF FLUID THROUGH THE DEVICE DURING INJECTION. THE BD PEGASUS IS AN INFUSION ONLY DEVICE AND IS NOT RATED FOR HIGH PRESSURE INJECTIONS.

Additional Manufacturer Narrative · 0

E.1. INITIAL REPORTER PHONE #: (B)(6). H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM THERE WAS LEAKAGE. THE FOLLOWING WAS RECEIVED BY THE INITIAL REPORTER: IN THE DEPARTMENT OF RADIOLOGY, WHEN A NURSE IS PUTTING AN INDWELLING NEEDLE ON A PATIENT FOR CONTRAST EXAMINATION, THE ISOLATION PLUG OF THE INDWELLING NEEDLE FALLS OFF, CAUSING BLOOD AND CONTRAST AGENT TO FLOW OUT OF THE BODY. IT DID NOT AFFECT THE MEDICAL STAFF, AND THE RESULTS OF THE PATIENT'S ANGIOGRAPHIC EXAMINATION WERE INCORRECT, SO A NEW EXAMINATION WAS REQUIRED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM THERE WAS LEAKAGE. THE FOLLOWING WAS RECEIVED BY THE INITIAL REPORTER: IN THE DEPARTMENT OF RADIOLOGY, WHEN A NURSE IS PUTTING AN INDWELLING NEEDLE ON A PATIENT FOR CONTRAST EXAMINATION, THE ISOLATION PLUG OF THE INDWELLING NEEDLE FALLS OFF, CAUSING BLOOD AND CONTRAST AGENT TO FLOW OUT OF THE BODY. IT DID NOT AFFECT THE MEDICAL STAFF, AND THE RESULTS OF THE PATIENT'S ANGIOGRAPHIC EXAMINATION WERE INCORRECT, SO A NEW EXAMINATION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2170793 BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 3080105

Patients

Seq Age Sex Outcome Treatment
1 Unknown