FDA Adverse Event
Other
Summary report: N
TERUMO CDI 500 BLOOD PARAMETER MONITOR
MDR report key: 1728941
·
Received June 14, 2010
Report
- Report Number
- 1828100-2010-00564
- Event Type
- Other
- Date Received
- June 14, 2010
- Date of Event
- May 21, 2010
- Report Date
- June 14, 2010
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP
- Product Code
- DRY
- PMA / PMN Number
- K972962
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
DURING PREPARATION FOR USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE USER REPORTED THE ARTERIAL MODULE SRS FAILURE ERROR CODE "F039", AND THE SERVICE TECHNICIAN OBSERVED A BROKEN CLIP. NO OTHER DETAILS REGARDING THE NATURE OF THE EVENT WERE PROVIDED. THERE WERE NO REPORTED ADVERSE CONSEQUENCES TO A PATIENT AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUMO CDI 500 BLOOD PARAMETER MONITOR | ON LINE BLOOD GAS MONITOR | DRY | TERUMO CARDIOVASCULAR SYSTEMS CORP | 500AHCT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |