HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36
Report
- Report Number
- 3005075853-2023-04782
- Event Type
- Malfunction
- Date Received
- July 10, 2023
- Date of Event
- May 30, 2023
- Report Date
- July 10, 2023
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- UDI-DI
- 10705036014621
- PMA / PMN Number
- K132612
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
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¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE. CONTINUED USAGE OF THE DAMAGED BLADE CAN RESULT IN A BROKEN BLADE. AS PART OF THE QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT/BATCH X9695Z, AND NO NON-CONFORMANCES WERE IDENTIFIED.
IT WAS REPORTED THAT DURING A GASTRECTOMY THE SCALPEL PASSED FIRST PRE-RUN TEST. AFTER WORKING SEVERAL MINUTES, THE SCALPEL WAS REQUIRED TO PRE-RUN BY THE SYSTEM AGAIN. THE SCALPEL COULD NOT WORK UNINTERRUPTED AND THEN THE DEVICE COULD NOT BE ACTIVATED DURING PROCEDURE. CHANGED TO ANOTHER DEVICE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 251298 | HARMONIC ACE PLUS 7 ADV HEMOSTASIS 36 | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | HARH36 | X9695Z | 10705036014621 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | HANDPIECE AND GENERATOR |