INFUSOMAT SPACE
Report
- Report Number
- 9610825-2023-00317
- Event Type
- Malfunction
- Date Received
- July 7, 2023
- Date of Event
- May 31, 2023
- Report Date
- October 5, 2023
- Manufacturer
- B.BRAUN MELSUNGEN AG
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN AG INTERNAL REPORT (B)(4) THE DEVICE HAS BEEN INVESTIGATED IN OUR SERVICE DEPARTMENT AT LABORATÓRIOS B.BRAUN S.A., SÃO GONÇALO, BRAZIL: THE INFUSION WAS STARTED BY THE USER AT 17:59:11, ON 05/30/2023. AND IT WAS COMPLETED BY THE USER AT 06:54:25 ON (B)(6) 2023. TOTAL INFUSION TIME 12H55MIN14. INFUSED VOLUME 565.24ML. RESULTS COMPATIBLE WITH THE INFUSION PERFORMED. TO VERIFY THE PERFORMANCE OF THE EQUIPMENT, WE PERFORM A FUNCTIONAL TEST USING THE INFORMATION PROVIDED BY THE USER. THERE IS NO EVIDENCE OF INFUSION ERROR IN THE HISTORY OF EQUIPMENT USE. THE RESULT OF THE FUNCTIONAL TEST INDICATES THAT THE EQUIPMENT IS WORKING ACCORDING TO THE ACCURACY SPECIFIED FOR IT. UNCONFIRMED TECHNICAL COMPLAINT. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN AG INTERNAL REPORT (B)(4). THE COMPLAINT IS UNDER EVALUATION. A FOLLOW-UP REPORT WILL BE PROVIDED AFTER THE EXAMINATION RESULTS ARE AVAILABLE. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.
AS REPORTED BY THE USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN BRAZIL: "OVER INFUSION / OPERATING UNIT". ACCORDING TO THE CUSTOMER: "CHILD USING DAILY NPT, INSTALLED AT 6 PM ON 05/30, PROGRAMMED 700ML IN 16 HOURS, PERFORMED A DOUBLE CHECK WITH THE MOTHER. NPT ENDED 3 HOURS BEFORE SCHEDULE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2015476 | INFUSOMAT SPACE | PUMP, INFUSION | FRN | B.BRAUN MELSUNGEN AG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |