Description of Event or Problem · 1
HYDRAULIC CENTURY-TUB LIFT BASE BROKE FROM THE FLOOR DUE TO CORROSION WITH A PATIENT OCCUPYING THE LIFT SEAT. PATIENT FELL WITH THE DEVICE AND FRACTURED FOUR (4) RIBSDEVICE NOT LABELED FOR SINGLE USE. PATIENT MEDICAL STATUS PRIOR TO EVENT: SATISFACTORY CONDITION. THERE WAS NOT MULTIPLE PATIENT INVOLVEMENT.DEVICE SERVICED IN ACCORDANCE WITH SERVICE SCHEDULE. DATE LAST SERVICED: 01-JUL-92. SERVICE PROVIDED BY: USER FACILITY BIOMEDICAL/BIOENGINEERING DEPARTMENT. SERVICE RECORDS NOT AVAILABLE.IMMINENT HAZARD TO PUBLIC HEALTH CLAIMED. DEVICE USED AS LABELED/INTENDED.DEVICE WAS EVALUATED AFTER THE EVENT. METHOD OF EVALUATION: ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED, VISUAL EXAMINATION. RESULTS OF EVALUATION: MATERIAL DEGRADATION/DETERIORATION. CONCLUSION: DEVICE FAILURE DIRECTLY CAUSED EVENT. CERTAINTY OF DEVICE AS CAUSE OF OR CONTRIBUTOR TO EVENT: YES. CORRECTIVE ACTIONS: DEVICE PERMANENTLY REMOVED FROM SERVICE, OTHER. THE DEVICE WAS NOT DESTROYED/DISPOSED OF.