FDA Adverse Event
Injury
Summary report: N
TUBE-STAT
MDR report key: 17270
·
Received October 28, 1994
Report
- Report Number
- MW1003851
- Event Type
- Injury
- Date Received
- October 28, 1994
- Date of Event
- October 25, 1994
- Report Date
- October 25, 1994
- Manufacturer
- XOMED TREACE, INC.
- Product Code
- BSR
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
ITEM IS PACKAGED WITH VINYL RING SLEEVE AROUND LIGHT BULB (TO PROTECT BULB). THE SLEEVE WAS NOT REMOVED BY THE OPERATOR PRIOR TO THE ORAL INTUBATION STYLET'S INTRODUCTION INTO THE LUMEN OF A 7.5 ID ENDOTRACHEAL TUBE. UPON REMOVAL OF THE INTUBATION STYLET THE VINYL RING SLEEVE BALLED UP IN THE ENDOTRACHEAL TUBE OCCLUDING THE LUMEN. THIS IS A POTENTIALLY LIFE-THREATENING PROBLEM. OF RELATED CONCERN IS THAT THIS SLEEVE MAY FALL OFF THE STYLET AND ENTER THE TRACHEOBRONCHIAL TREE. THE VINYL RING SLEEVE (COLLAR) AS PACKAGED IS COLORLESS AND EASILY OVERLOOKED. IT MAY BE REASONABLE TO SUGGEST TO THE MFR TO COLOR-CODE THE PROTECTIVE SLEEVE, OR TO PACKAGE THE INSTRUMENT SO THAT THE LIGHT BULB IS CUSHIONED INTO THE PACKAGING MATERIAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TUBE-STAT | ORAL INTUBATION STYLET | BSR | XOMED TREACE, INC. | 83-30400 | 81166 00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Required Intervention |