FDA Adverse Event Injury Summary report: N

TUBE-STAT

MDR report key: 17270 · Received October 28, 1994

Report

Report Number
MW1003851
Event Type
Injury
Date Received
October 28, 1994
Date of Event
October 25, 1994
Report Date
October 25, 1994
Manufacturer
XOMED TREACE, INC.
Product Code
BSR
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

ITEM IS PACKAGED WITH VINYL RING SLEEVE AROUND LIGHT BULB (TO PROTECT BULB). THE SLEEVE WAS NOT REMOVED BY THE OPERATOR PRIOR TO THE ORAL INTUBATION STYLET'S INTRODUCTION INTO THE LUMEN OF A 7.5 ID ENDOTRACHEAL TUBE. UPON REMOVAL OF THE INTUBATION STYLET THE VINYL RING SLEEVE BALLED UP IN THE ENDOTRACHEAL TUBE OCCLUDING THE LUMEN. THIS IS A POTENTIALLY LIFE-THREATENING PROBLEM. OF RELATED CONCERN IS THAT THIS SLEEVE MAY FALL OFF THE STYLET AND ENTER THE TRACHEOBRONCHIAL TREE. THE VINYL RING SLEEVE (COLLAR) AS PACKAGED IS COLORLESS AND EASILY OVERLOOKED. IT MAY BE REASONABLE TO SUGGEST TO THE MFR TO COLOR-CODE THE PROTECTIVE SLEEVE, OR TO PACKAGE THE INSTRUMENT SO THAT THE LIGHT BULB IS CUSHIONED INTO THE PACKAGING MATERIAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TUBE-STAT ORAL INTUBATION STYLET BSR XOMED TREACE, INC. 83-30400 81166 00

Patients

Seq Age Sex Outcome Treatment
1 * Required Intervention