FDA Adverse Event Injury Summary report: N

KIT IMPLANTABLE SLIM TIP LEAD, 50CM

MDR report key: 17263404 · Received July 5, 2023

Report

Report Number
1627487-2023-03119
Event Type
Injury
Date Received
July 5, 2023
Date of Event
June 10, 2023
Report Date
October 6, 2023
Manufacturer
ABBOTT MEDICAL
Product Code
PMP
PMA / PMN Number
P150004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT IS ESTIMATED. THE ALLEGATION IS AGAINST 1 OF 4 LEADS; HOWEVER, IT IS UNKNOWN WHICH LEAD, THEREFORE, ALL POTENTIAL COMPONENTS ARE BEING LISTED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT INCLUDE: COMMON DEVICE NAME: SLIM TIP LEAD, MODEL: MN10450-50A, SERIAL: (B)(6), UDI: (B)(4), BATCH: 8081985; COMMON DEVICE NAME: SLIM TIP LEAD, MODEL: MN20450-50A, SERIAL: (B)(6), UDI: (B)(4), BATCH: 8145749; COMMON DEVICE NAME: SLIM TIP LEAD, MODEL: MN20450-50A, SERIAL: (B)(6), UDI: (B)(4), BATCH: 8145749.

Additional Manufacturer Narrative · 0

THE REPORT OF INEFFECTIVE STIMULATION WAS CONFIRMED. ANALYSIS OF THE LEAD A FOUND THAT IT WAS RETURNED IN TWO SEGMENTS AS A RESULT OF THE EXPLANT PROCEDURE. MICROSCOPIC INSPECTION OF THE STIMULATION END SEGMENT FOUND A KINK ON THE OUTER TUBING WHERE ALL INTERNAL WIRES WERE BROKEN. THESE FRACTURES ARE CONSISTENT WITH AN OVERSTRESS CONDITION OR SUDDEN EVENT THE LEAD WAS SUBJECTED TO WHILE STILL IN VIVO. NO EVALUATION FORM USED.

Additional Manufacturer Narrative · 0

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION DURING WHICH THE SYSTEM WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1930929 KIT IMPLANTABLE SLIM TIP LEAD, 50CM DRG LEAD PMP ABBOTT MEDICAL MN20450-50A 8145749

Patients

Seq Age Sex Outcome Treatment
1 Female Other DRG IPG| DRG LEAD (X3)