KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Report
- Report Number
- 1627487-2023-03119
- Event Type
- Injury
- Date Received
- July 5, 2023
- Date of Event
- June 10, 2023
- Report Date
- October 6, 2023
- Manufacturer
- ABBOTT MEDICAL
- Product Code
- PMP
- PMA / PMN Number
- P150004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DATE OF EVENT IS ESTIMATED. THE ALLEGATION IS AGAINST 1 OF 4 LEADS; HOWEVER, IT IS UNKNOWN WHICH LEAD, THEREFORE, ALL POTENTIAL COMPONENTS ARE BEING LISTED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT INCLUDE: COMMON DEVICE NAME: SLIM TIP LEAD, MODEL: MN10450-50A, SERIAL: (B)(6), UDI: (B)(4), BATCH: 8081985; COMMON DEVICE NAME: SLIM TIP LEAD, MODEL: MN20450-50A, SERIAL: (B)(6), UDI: (B)(4), BATCH: 8145749; COMMON DEVICE NAME: SLIM TIP LEAD, MODEL: MN20450-50A, SERIAL: (B)(6), UDI: (B)(4), BATCH: 8145749.
THE REPORT OF INEFFECTIVE STIMULATION WAS CONFIRMED. ANALYSIS OF THE LEAD A FOUND THAT IT WAS RETURNED IN TWO SEGMENTS AS A RESULT OF THE EXPLANT PROCEDURE. MICROSCOPIC INSPECTION OF THE STIMULATION END SEGMENT FOUND A KINK ON THE OUTER TUBING WHERE ALL INTERNAL WIRES WERE BROKEN. THESE FRACTURES ARE CONSISTENT WITH AN OVERSTRESS CONDITION OR SUDDEN EVENT THE LEAD WAS SUBJECTED TO WHILE STILL IN VIVO. NO EVALUATION FORM USED.
THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED.
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INEFFECTIVE THERAPY. AS A RESULT, THE PATIENT UNDERWENT SURGICAL INTERVENTION DURING WHICH THE SYSTEM WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1930929 | KIT IMPLANTABLE SLIM TIP LEAD, 50CM | DRG LEAD | PMP | ABBOTT MEDICAL | MN20450-50A | 8145749 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Other | DRG IPG| DRG LEAD (X3) |