CARDIOSAVE HYBRID, TYPE B PLUG
Report
- Report Number
- 2249723-2023-03081
- Event Type
- Malfunction
- Date Received
- July 5, 2023
- Date of Event
- June 26, 2023
- Report Date
- March 27, 2024
- Manufacturer
- DATASCOPE CORP. - MAHWAH
- Product Code
- DSP
- UDI-DI
- 10607567108391
- PMA / PMN Number
- K112372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.
UPDATED FIELDS: B4, D9, G3, G6, H2, H3, H4, H6(TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS), H10. A GETINGE FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE UNIT. THE (FSE) REPLACED THE (D441-00-0194 COVER, CONSOLE), (D997-00-0587 ASSY,SLIDE HANDLE), (D997-00-0588 ASSEMBLY, WHEELS.) , (D361-00-0197 STANDOFF, PIVOT) , (D125-00-0028 CABLE TIE, 5/16 NYLON P-CLIP) WHICH HAD BEEN DAMAGED BY THE CUSTOMER. THE UNIT PASSED ALL FUNCTIONAL AND SAFETY TESTS AND WAS RETURNED TO CUSTOMER.
IT WAS REPORTED THAT THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) PORTABLE PORTION HAS A BIG DENT. INTERNAL PRESSURE HOSE IS BEING COMPRESSED. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
N/A.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1254681 | CARDIOSAVE HYBRID, TYPE B PLUG | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. - MAHWAH | 0998-00-0800-53 | 10607567108391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |