FDA Adverse Event Malfunction Summary report: N

CARDIOSAVE HYBRID, TYPE B PLUG

MDR report key: 17262452 · Received July 5, 2023

Report

Report Number
2249723-2023-03081
Event Type
Malfunction
Date Received
July 5, 2023
Date of Event
June 26, 2023
Report Date
March 27, 2024
Manufacturer
DATASCOPE CORP. - MAHWAH
Product Code
DSP
UDI-DI
10607567108391
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

UPDATED FIELDS: B4, D9, G3, G6, H2, H3, H4, H6(TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS), H10. A GETINGE FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE UNIT. THE (FSE) REPLACED THE (D441-00-0194 COVER, CONSOLE), (D997-00-0587 ASSY,SLIDE HANDLE), (D997-00-0588 ASSEMBLY, WHEELS.) , (D361-00-0197 STANDOFF, PIVOT) , (D125-00-0028 CABLE TIE, 5/16 NYLON P-CLIP) WHICH HAD BEEN DAMAGED BY THE CUSTOMER. THE UNIT PASSED ALL FUNCTIONAL AND SAFETY TESTS AND WAS RETURNED TO CUSTOMER.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) PORTABLE PORTION HAS A BIG DENT. INTERNAL PRESSURE HOSE IS BEING COMPRESSED. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

Description of Event or Problem · 0

N/A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1254681 CARDIOSAVE HYBRID, TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. - MAHWAH 0998-00-0800-53 10607567108391

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown