FDA Adverse Event Malfunction Summary report: N

BD POSIFLUSH¿ NORMAL SALINE SYRINGE

MDR report key: 17258791 · Received July 5, 2023

Report

Report Number
1911916-2023-00455
Event Type
Malfunction
Date Received
July 5, 2023
Date of Event
June 8, 2023
Report Date
June 14, 2023
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
NGT
PMA / PMN Number
K161552
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E.1 INITIAL REPORTER PHONE #: (B)(6). H.6 INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 306595 AND LOT NUMBER 2011656. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT 3 OF THE BD POSIFLUSH¿ NORMAL SALINE SYRINGE EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: DURING THE USE OF THE PRODUCT, IT WAS FOUND THAT THE CONNECTION BETWEEN THE NEEDLELESS CONNECTOR (MP1000 CHINA) AND THE FLUSH LEAKED, AND IT WAS IMPOSSIBLE TO CONFIRM WHICH PRODUCT CAUSED THE LEAKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1787117 BD POSIFLUSH¿ NORMAL SALINE SYRINGE SALINE, VASCULAR ACCESS FLUSH NGT BD MEDICAL (BD WEST) MEDICAL SURGICAL 2011656

Patients

Seq Age Sex Outcome Treatment
1 Unknown