PUMP MMT-1712KL 640G V4.10 BK SF MG
Report
- Report Number
- 2032227-2023-230589
- Event Type
- Malfunction
- Date Received
- July 5, 2023
- Date of Event
- April 20, 2023
- Report Date
- July 4, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED OUTSIDE THE UNITED STATES. P-CAP LOCKED IN PLACE PROPERLY DURING TESTING. AFTER BATTERY INSTALLATION UNIT GAVE AN UNEXPECTED FLASHING WHITE / BLANK DISPLAY FOLLOWED BY AN UNEXPECTED INTERMITTENT BEEP ALARM. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO DISPLAY ANOMALY. PROCEEDED BY CUTTING UNIT OPEN AND PERFORM VISUAL INSPECTION OF CONNECTORS AND ELECTRONIC ASSEMBLY. PER VISUAL INSPECTION IT WAS DETERMINED THE INGRESS OF WATER CORRODING AND CAUSING ELECTRICAL SHORTING ON ELECTRONIC ASSEMBLY, MOTOR ASSEMBLY AND KEYPAD ASSEMBLY. UNABLE TO DOWNLOAD HISTORY FILES AND TRACES USING THUS DUE TO DISPLAY ANOMALY. FORCE SENSOR ZERO OFFSET WITHIN SPEC (23.1MV). UNIT ALSO RECEIVED WITH CRACKED CASE (BATTERY TUBE) AND CRACKED BATTERY TUBE THREADS. IN CONCLUSION UNIT RECEIVED WITH UNEXPECTED BLANK WHITE FLASHING SCREEN DUE TO CORRODED ELECTRONIC ASSEMBLY. ISOLATE TO ELECTRONIC ASSEMBLIES. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED CRACK ALONG THE BATTERY COMPARTMENT. TROUBLESHOOTING WAS NOT PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WAS RETURNED FOR FAILURE ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418876 | PUMP MMT-1712KL 640G V4.10 BK SF MG | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1712KL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |