INCLUSIVE TAPERED IMPLANT 3.7 MMD X 10 MML X 3.5 MMP
Report
- Report Number
- 3011649314-2023-00420
- Event Type
- Injury
- Date Received
- July 5, 2023
- Date of Event
- August 10, 2022
- Report Date
- June 13, 2025
- Manufacturer
- PRISMATIK DENTALCRAFT, INC.
- Product Code
- DZE
- PMA / PMN Number
- K121406
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THE DEVICE HAS NOT BEEN RETURNED. IF/WHEN THERE IS MORE INFORMATION PROVIDED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. SECTION A4: THE PATIENTS WEIGHT IS NOT PROVIDED WHEN ASKED. SECTION A5/6: THE PATIENTS WEIGHT IS NOT PROVIDED WHEN ASKED. SECTION B3: THIS INFORMATION WAS NOT PROVIDED WHEN ASKED. SECTION D6B: THIS INFORMMATION WAS NOT PROVIDED WHEN ASKED.
THE DEVICE WAS RETURNED, THE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR INCLUSIVE TAPERED IMPLANT LOT# 6111905 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: A REVIEW OF STOCK PRODUCT WAS PERFORMED FOR INCLUSIVE TAPERED IMPLANT LOT #6111905 AND FOUND NO ADDITIONAL PRODUCT IN STOCK TO REVIEW. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE AN INCLUSIVE TAPERED IMPLANT 3.7 MMD X 10.0 MML X 3.5 MMP (70-1070-IMP0006) USING RADIOGRAPHIC TEMPLATE. THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. BONE DEBRIS WAS OBSERVED IN THE THREADING OF THE IMPLANT. ROOT CAUSE: "LOSS OF OSSEOINTEGRATION" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LOSS OF OSSEOINTEGRATION IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE THE LOSS OF PRIMARY STABILITY AT THE OSTEOTOMY SITE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU 4989 REV 5.0 (INCLUSIVE TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "INCLUSIVE TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE" IFU 4989 REV 5.0 (INCLUSIVE TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE PRECAUTION SECTION SURGICAL PROCEDURES: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO. SINCE IMPLANT COMPONENTS AND THEIR INSTRUMENTS ARE VERY SMALL, PRECAUTIONS SHOULD BE TAKEN TO ENSURE THAT THEY ARE NOT SWALLOWED OR ASPIRATED BY THE PATIENT. PRIOR TO SURGERY, ENSURE THAT THE NEEDED COMPONENTS, INSTRUMENTS AND ANCILLARY MATERIALS ARE COMPLETE, FUNCTIONAL AND AVAILABLE IN THE CORRECT QUANTITIES" IFU 4989 REV 5.0 (INCLUSIVE TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: " THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." IFU 4989 REV 5.0 (INCLUSIVE TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE, AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION."
ADDITIONAL INFORMATION: A4, A6, B1, H6. CORRECTED INFORMATION: A1, A2, B5, D4, E1, G1, H1. E1: IN THE INITIAL REPORT, THE ESTABLISHMENT NAME WAS REPORTED AS (B)(6), HOWEVER, THIS INFORMATION WAS NOT PROVIDED. CAPA CA-00016. MANUFACTURER REFERENCE: (B)(4). REPORTS FOR THE ADDITIONAL REPORTED DEVICES/EVENTS ARE FOUND IN THE FOLLOWING MDRS: 3011649314-2023-00418, 3011649314-2023-00419, 3011649314-2023-00421.
IT WAS REPORTED THAT THE INCLUSIVE TAPERED IMPLANT FAILED. THE PATIENT BONE GRADE IS NOTED AS GRADE III WITH THE IMPLANT PLACED ON (B)(6( 2022. THE RETURNED ON AN UNKNOWN DATE. UPON EXAM THE PROVIDER NOTED TO HAVE LOSS INTEGRATION AND WAS REMOVED.
IT WAS REPORTED THAT THE INCLUSIVE TAPERED IMPLANT FAILED. THE PATIENT BONE GRADE IS NOTED AS GRADE III. THE IMPLANT WAS PLACED ON (B)(6) 2022. THE PATIENT PRESENTED WITH THE ISSUE ON (B)(6) 2022. UPON EXAM THE PROVIDER NOTED TO HAVE LOSS INTEGRATION AND WAS REMOVED.
IT WAS REPORTED THAT THE INCLUSIVE TAPERED IMPLANTS FAILED. THE PATIENT BONE GRADE IS NOTED AS GRADE III. THE PATIENT PRESENTED FOR IMPLANT PLACEMENT ON TEETH POSITIONS #20, 22, 26 AND 28 ON (B)(6) 2022. THE PATIENT PRESENTED ON (B)(6) 2022 FOR FOLLOW-UP WITH COMPLIANT OF LOOSE DENTURES. UPON EXAM THE PROVIDER NOTED THE IMPLANT SITES WERE INFECTED AND THAT THE IMPLANTS LOSS OSSEOINTEGRATION. THE IMPLANTS WERE REMOVED AND NOT REPLACED. ADDITIONAL INFORMATION REPORTED THE DATE OF EXPLANT AS (B)(6) 2022. GRAFTING WAS PERFORMED, THERE WAS NO PATIENT INJURY AND THE PATIENTS SYMPTOMS WERE RELIEVED AFTER IMPLANT REMOVAL. THE PATIENT STATUS WAS REPORTED AS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1775182 | INCLUSIVE TAPERED IMPLANT 3.7 MMD X 10 MML X 3.5 MMP | INCLUSIVE TAPERED IMPLANT | DZE | PRISMATIK DENTALCRAFT, INC. | 70-1070-IMP0006 | 6111905 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Female | Required Intervention |