FDA Adverse Event
Other
Summary report: N
PUMP
MDR report key: 17248305
·
Received June 30, 2023
Report
- Report Number
- MW5119070
- Event Type
- Other
- Date Received
- June 30, 2023
- Manufacturer
- UNK
- Product Code
- FRN
- Report Source
- Voluntary report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- *
Narratives
Description of Event or Problem · 0
NITROGLYCERIN 200 MCG/ML WAS PROGRAMMED TO RUN AT 10ML/HR INSTEAD OF THE ORDERED 10 MCG/MIN. SO THE PATIENT RECEIVED 33 MCG/MIN INSTEAD OF THE ORDERED 20MCG/MIN. CURRENT GUARDRAILS HARD MAX IS SET AT 45 MCG/MIN. THIS HAS ALSO OCCURRED WITH FUROSEMIDE DRIPS 2MG/ML. PUMP SET TO RUN AT 10ML/HR(5MG/HR) INSTEAD OF DOSE OF 10MG/HR. (B)(6), PHONE: (B)(6), EMAIL: (B)(6). SUBMISSION ID: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2015264 | PUMP | PUMP, INFUSION | FRN | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | FUROSEDMIDE| NIGROGLYCERIN INJECTION 200 MCG/ML |