ENDOWRIST SP
Report
- Report Number
- 2955842-2023-16744
- Event Type
- Malfunction
- Date Received
- June 30, 2023
- Date of Event
- May 3, 2023
- Report Date
- May 31, 2023
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874114322
- PMA / PMN Number
- K182371
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- 003
Narratives
INTUITIVE SURGICAL, INC. (ISI) FAILURE ANALYSIS FURTHER INVESTIGATED AND FOUND THE INSTRUMENT TO HAVE A BROKEN MOLDED INSULATOR. THE MOLDED INSULATOR BROKE AND NONE OF THE PIECES WERE RETURNED WITH THE INSTRUMENT. THE MOLDED INSULATOR SECURES THE GRIP SUBASSEMBLY TO THE OVERALL GRIP SET, AND IF IT BREAKS COMPLETELY THE GRIP CAN BECOME DISLODGED. THIS INSTRUMENT ALSO APPEARS TO HAVE A BROKEN BIPOLAR WIRE AT THE DISTAL END AND MAY HAVE BROKEN DURING THE SAME EVENT WHERE THE MOLDED INSULATOR BROKE, AND THE GRIP BECAME DISLODGED.
INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE MARYLAND BIPOLAR FORCEPS INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A BROKEN GRIP AT THE GRIP BASE. A PIECE APPROXIMATELY 0.168" X 0.598" WAS FOUND TO BE BROKEN OFF. THE BROKEN PIECE WAS NOT RETURNED. THE COMPLAINT REGARDING THE INSTRUMENT BEING BROKEN WAS CONFIRMED BY FAILURE ANALYSIS.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED SURGICAL PROCEDURE THAT THE MARYLAND BIPOLAR FORCEPS (MBF) INSTRUMENT HAD ONLY ONE USE LEFT AND WAS BROKEN. THERE WERE NO REPORTS OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 26327 | ENDOWRIST SP | MARYLAND BIPOLAR FORCEPS | NAY | INTUITIVE SURGICAL, INC | 430010-57 | U10220908 0010 | 00886874114322 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |