FDA Adverse Event Malfunction Summary report: N

ENDOWRIST SP

MDR report key: 17244022 · Received June 30, 2023

Report

Report Number
2955842-2023-16744
Event Type
Malfunction
Date Received
June 30, 2023
Date of Event
May 3, 2023
Report Date
May 31, 2023
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874114322
PMA / PMN Number
K182371
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) FAILURE ANALYSIS FURTHER INVESTIGATED AND FOUND THE INSTRUMENT TO HAVE A BROKEN MOLDED INSULATOR. THE MOLDED INSULATOR BROKE AND NONE OF THE PIECES WERE RETURNED WITH THE INSTRUMENT. THE MOLDED INSULATOR SECURES THE GRIP SUBASSEMBLY TO THE OVERALL GRIP SET, AND IF IT BREAKS COMPLETELY THE GRIP CAN BECOME DISLODGED. THIS INSTRUMENT ALSO APPEARS TO HAVE A BROKEN BIPOLAR WIRE AT THE DISTAL END AND MAY HAVE BROKEN DURING THE SAME EVENT WHERE THE MOLDED INSULATOR BROKE, AND THE GRIP BECAME DISLODGED.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE MARYLAND BIPOLAR FORCEPS INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A BROKEN GRIP AT THE GRIP BASE. A PIECE APPROXIMATELY 0.168" X 0.598" WAS FOUND TO BE BROKEN OFF. THE BROKEN PIECE WAS NOT RETURNED. THE COMPLAINT REGARDING THE INSTRUMENT BEING BROKEN WAS CONFIRMED BY FAILURE ANALYSIS.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED SURGICAL PROCEDURE THAT THE MARYLAND BIPOLAR FORCEPS (MBF) INSTRUMENT HAD ONLY ONE USE LEFT AND WAS BROKEN. THERE WERE NO REPORTS OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
26327 ENDOWRIST SP MARYLAND BIPOLAR FORCEPS NAY INTUITIVE SURGICAL, INC 430010-57 U10220908 0010 00886874114322

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES