FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE

MDR report key: 17242053 · Received June 30, 2023

Report

Report Number
1024879-2023-00429
Event Type
Malfunction
Date Received
June 30, 2023
Date of Event
June 19, 2023
Report Date
August 2, 2023
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
50382903686075
PMA / PMN Number
K982541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 20 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING, EACH HAVING THEIR ECLIPSE SHIELDS ACTIVATED, AND NO ISSUES WERE OBSERVED RELATING TO CANNULA BREAKING OFF AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE CANNULA BREAKING OFF. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THAT THE CANNULA BREAKS OFF OR PULL OUT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS THAT NEEDLE HAS BEEN BREAKING FOR CAT 368607 LOT 3052518.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THAT THE CANNULA BREAKS OFF OR PULL OUT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS THAT NEEDLE HAS BEEN BREAKING FOR CAT 368607 LOT 3052518.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
22652 BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON, DICKINSON & CO., (BD) 368607 3052518 50382903686075

Patients

Seq Age Sex Outcome Treatment
1 Unknown