BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE
Report
- Report Number
- 1024879-2023-00429
- Event Type
- Malfunction
- Date Received
- June 30, 2023
- Date of Event
- June 19, 2023
- Report Date
- August 2, 2023
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- FMI
- UDI-DI
- 50382903686075
- PMA / PMN Number
- K982541
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 20 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING, EACH HAVING THEIR ECLIPSE SHIELDS ACTIVATED, AND NO ISSUES WERE OBSERVED RELATING TO CANNULA BREAKING OFF AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE CANNULA BREAKING OFF. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.
H.3. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THAT THE CANNULA BREAKS OFF OR PULL OUT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS THAT NEEDLE HAS BEEN BREAKING FOR CAT 368607 LOT 3052518.
IT WAS REPORTED THAT WHILE USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THAT THE CANNULA BREAKS OFF OR PULL OUT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS THAT NEEDLE HAS BEEN BREAKING FOR CAT 368607 LOT 3052518.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 22652 | BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BECTON, DICKINSON & CO., (BD) | 368607 | 3052518 | 50382903686075 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |