FDA Adverse Event Other Summary report: N

LEVER LOCK ADAPTOR

MDR report key: 172403 · Received June 10, 1998

Report

Report Number
172403
Event Type
Other
Date Received
June 10, 1998
Date of Event
May 20, 1998
Report Date
June 4, 1998
Manufacturer
BECTON DICKINSON
Product Code
FPA
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
SD, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

INTRAVENOUS CHEMOTHERAPY INITIATED ON PT (05/20/1998) THROUGH A PERIPHERALLY INSERTED CENTRAL CATHETER LINE PRIOR TO DISCHARGE (05/21/1998). HOME HEALTH NURSE IDENTIFIED THE INFUSION WAS LEAKING WHEN SHE SAW THE CLIENT ON 05/22/1998. SHE IMMEDIATELY NOTIFIED FACILITY AND CLEANED UP THE SPILL PER PROTOCOL USING A CHEMO SPILL KIT. THE HUSBAND STATED HE HAD NOTED BLOOD BACKING UP IN PERIPHERALLY INSERTED CENTRAL CATHETER LINE AT APPROXIMATELY 1630 ON 05/21/1998 AFTER BRINGING HIS WIFE HOME. IT WAS DETERMINED THAT APPROXIMATELY 44 ML OF CHEMO AGENT HAD LEAKED OUT (2 ML/HOUR X 22 HOURS). WHEN THE PERIPHERALLY INSERTED CENTRAL CATHETER LINE WAS ATTACHED TO THE INTRAVENOUS PUMP TUBING, AN INCORRECT CONNECTOR PIECE WAS USED. THIS IS THE SITE WHERE THE LEAK OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LEVER LOCK ADAPTOR IV TUBING CONNECTOR FPA BECTON DICKINSON UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other